ISO/IEC 17025 in Analytical Laboratories

1. Management Summary
ISO/IEC 17025 is a quality standard for testing and calibration
laboratories. The current release was published in 2005. There are
two main sections in ISO/IEC 17025 - Management Requirements and
Technical Requirements. Management requirements are related to the
operation and effectiveness of the quality management system within
the laboratory and has similar requirements to ISO 9001. Technical
requirements address the competence of staff, testing methodology,
equipment and quality and reporting of test and calibration results.
The standard is the basis for accreditation from an accreditation
body.
Implementing ISO/IEC 17025 has benefits for laboratories but there
are also additional work and costs required.
Main benefits are:
- Having ISO/IEC 17025 accreditation status
will get direct access to more contracts for testing and/or
calibration. Some public and private organizations only give
contracts to accredited laboratories.
- Having ISO/IEC 17025 accreditation status
will improve the reputation and image of the laboratory. This
will also help to get more contracts from organizations that
don’t mandate accreditation but give preference to accredited
laboratories in competitive situations.
- When correctly implemented, the quality
system can help to continually improve the quality of data and
effectiveness of the laboratory.
- ISO/IEC 17025 is the basis for most other
quality systems related to laboratories, for example, Good
Manufacturing Practices and Good Laboratory Practices.
Analytical testing laboratories seeking ISO
17025 will be impacted in a couple of areas. The main difference
between formal accreditation and 'just' good analytical practices is
the amount of documentation to be developed. There is no doubt that
any good analytical laboratory uses qualified analysts, and for
performing tests, checks the performance of equipment used for
testing and validates analytical methods. However, many times the
outcome of the tests is not fully documented. ISO 17025
accreditation requires formal documentation for about everything.
It's similar to operating in a regulated environment; ‘what is not
documented is a rumor’, assessors consider it as 'not being done'.
The overall impact on analytical laboratories can be best
illustrated on the entire sample/data workflow in a laboratory.
Figure 1 shows a typical workflow of samples
and test data along with the ISO 17025 requirements on individual
steps and the laboratory.

Figure 1: ISO/IEC 17025 Requirements for
Analytical Laboratories
- Sampling should be performed according to
a sampling plan and all sample details should be documented.
- Samples should be uniquely identified and
the sample integrity should be protected during transport and
storage.
- Quality of test results should be
monitored.
- The test report should not only include
test results, but also an estimation of the overall measurement
uncertainty. It should also include either detailed information
on the sample and test condition, or a link to a reference
document.
- Records should be well maintained to
ensure data integrity and availability.
There are also some requirements that impact
more than one workflow step:
- All analytical methods and procedures
should be validated. This includes methods and procedures for
sampling, testing and data evaluation.
- Equipment used for sampling and testing
should be calibrated, tested and well maintained. Material such
as calibration standards should be qualified and traceable to SI
units or to certified reference material.
- Nonconforming test results should be
documented and controlled.
- People should be qualified for the
assigned task, for example, through education, experience or
training.
- Environmental conditions such as
temperature, humidity and electromagnetic interference should be
monitored and controlled.
- All routine tasks should follow written
procedures.
- There are additional requirements that
not only impact the sample analysis but also the entire
organization of the laboratory.
- Specific documents should be developed
and maintained, for example, policies and a quality plan.
- Known current problems should not only be
corrected but a preventive action plan should be developed to
avoid reoccurrence of the same or similar problems.
- All complaints from clients should be
formally followed up.
- There should be a formal program to
manage suppliers, service providers and subcontractors.
- The organizational structure should be
such that there are no conflicting interests that could impact
quality.
- Compliance with ISO/IEC 17025 and
internal procedures should be assessed during regular internal
audits.
Key steps towards accreditation are:
- Management defines a project owner.
- The project owner studies details of the
standard and supporting literature and other relevant
information.
- The project owner defines the preliminary
scope of accreditation and prepares a list with requirements
together with the laboratory.
- The project owner performs a gap analysis
with the help of the laboratory professionals to see the
difference between the requirements and what is currently
implemented in the laboratory.
- Based on the outcome of the gap analysis
the project owner together with laboratory, financing and
documentation professionals and external consultants estimates
the costs for accreditation.
- Estimated costs are presented to
management together with incremental opportunities.
- Management decides to go for the project.
- The project owner leads implementation
steps.
Companies have to continuously deliver
high-quality products and/or services if they want to be successful
in the marketplace in the long term. Quality improvement has become
a key national and international business strategy. Most companies
are using quality systems as a method of assuring the consistent
conformity of products or services to a defined set of standards or
customer expectations.
Quality Systems
Several Quality System Standards were
developed in various countries in the 1960's and 1970's. The
MIL-Q-9858A in the United States in 1963 and the BS 5750 in the
United Kingdom in 1979 are probably the most important ones. The ISO
9000 series of quality standards were established in 1987 for
implementing and maintaining a quality system which is
internationally accepted and can be used as a criterion for third
party quality assessment.
ISO/IEC 17025 - Laboratory Quality System
Laboratories play an important role in the
quality systems of the companies. The ISO/IEC 17025 (1) can be used
as a standard to develop and establish a quality system in the
laboratory and for assessment by their clients or third parties. The
standard is also being used as a criterion for laboratory
accreditation.
The first edition (1999) of the International
Standard "General Requirements for the Competence of Testing and
Calibration Laboratories" was produced as a result of extensive
experience in the implementation of ISO/IEC Guide 25 and EN 45001,
both of which it replaced. It contains all the requirements that
testing and calibration laboratories have to meet if they wish to
demonstrate that they operate a management system, are technically
competent and are able to generate technically valid results.
Management requirements of the first edition
referred to ISO 9001:1994 and ISO 9002:1994. These standards have
been superseded by ISO 9001:2000, which made an alignment of ISO/IEC
17025 necessary. In the second edition of ISO 17025, released in
2005, clauses were amended or added only when considered necessary
in the light of ISO 9001:2000. Testing and calibration laboratories
that comply with this International Standard will therefore also
operate in accordance with ISO 9001. Accreditation bodies that
recognize the competence of testing and calibration laboratories use
this International Standard as the basis for their accreditation.
ISO/IEC 17025 is divided into five chapters, two annexes and one
bibliography section:
- Chapter 1: Scope
- The standard covers the technical
activities of a laboratory and the management and organizational
aspects to perform the technical activities in a competent way.
- Chapter 2: Normative References
- Chapter 3: Terms and Definitions
- Chapter 4: Management Requirements
Most of the requirements are similar to those specified in the
ISO Standard 9001:2000
- Chapter 5: Technical Requirements
Most of the requirements come from the ISO Guide 25.
- Annex A: Cross References to ISO
9001:2000
- Annex B: Guidelines for Establishing
Applications for Specific Fields
- Bibliography
Most important are chapters 4 and 5 on
management and technical requirements. Besides official requirements
these chapters also include notes with further explanations and
recommendations.
Scope and Contents of this Tutorial
Implementing a quality system such as ISO/IEC
17025 has an impact on a laboratories organization and operation.
This tutorial will discuss some of the specific requirements along
with their implications for testing laboratories.
This tutorial is especially useful for chemical analytical
laboratories that seek accreditation according to an internationally
recognized standard. Examples are food testing, environmental
testing, chemical testing, clinical testing, pharmaceutical testing
and other testing laboratories.
The tutorial will guide laboratory personnel and managers and QA
managers and staff through the entire process of ISO/IEC 17025
accreditation.
It also helps laboratories that have to work under different quality
systems to efficiently set up procedures for compliance with all
requirements.
It covers:
- Management requirements.
- Technical requirements.
- • Recommendations for Implementation.
- Steps toward ISO/IEC 17025 Accreditation.
Documentation
- Internal and External Audits
- Implementing multiple quality systems
The tutorial and its reference material should
give a good understanding of why 17025 is important, what the
requirements are and what are the key points for implementation. It
also helps to improve the overall quality of analytical results and
at the same time, to improve the recognition of the laboratory and
its employees.
The tutorial is not a substitute for the
standard itself. It does not list all requirements but rather it
focuses on the most important ones and the ones that need specific
attention according to the opinion of the author. The tutorial also
does not include tools such as example quality manual, operating
procedures and all the templates that would help to quickly
implement ISO/IEC 17025. For this we recommend looking for special
packages that are available from service providers, for example, the
ISO 17025 Accreditation Package from Labcompliance (2).

Management requirements are related to the
operation and effectiveness of the quality management system within
the laboratory and has similar requirements to ISO 9001. This part
is divided into fifteen chapters.
Organization
This chapter ensures that the roles and
responsibilities of the laboratory, the management and key personnel
are defined.
Key points are:
- An organizational structure and
responsibilities and tasks of management and staff should be
defined.
- The organizational structure should be
such that departments having conflicting interests do not
adversely influence the laboratory´s quality of work. Examples
are commercial marketing or financing.
- A quality assurance manager should be
appointed.
- All personnel should be free from any
commercial and financial pressure that may adversely impact the
quality of calibration and test results.
Management System
- This chapter should ensure that a
management system is implemented, maintained and continually
improved.
- Key points are:
- There should be policies, standard
procedures and work instructions to ensure the quality of test
results.
- There should be a quality manual with
policy statements that are issued and communicated by top level
management.
- The effectiveness of the management
system should be continually improved.
Document Control
Individual paragraphs in this chapter should
ensure that all documents related to the management system are
uniquely identified and created, approved, issued and changed
following documented procedures.
Key points are:
- All official documents should be
authorized and controlled.
- Documents should be regularly reviewed
and updated if necessary. The review frequency depends on the
document itself. Typical review cycles are between one and three
years.
- Changes to documents should follow the
same review process as for development of initial documents.
- Review of Requests, Tenders and Contracts
- This chapter should ensure that
requirements of requests, tenders and contracts are well
defined, reviewed, understood and documented.
- Key points are:
- The review by the laboratory supervisors
should ensure that the laboratory has technical capability and
the resources to meet the requirements.
- Changes in a contract should follow the
same process as the initial contract.
Subcontracting of Tests and Calibrations
- This chapter should ensure that tests and
calibrations subcontracted to 3rd parties are performed
according to the same quality standards as if they were done in
the subcontracting laboratory.
Key points are:
- The competence of the subcontracted party
should be ensured, for example, through a documented quality
system, e.g., ISO 17025.
- The subcontracting laboratory is
responsible to the customer for the subcontractor’s work.
Purchasing Services and Supplies
This chapter should ensure that services and
supplies delivered by 3rd parties do not adversely impact the
quality and effectiveness of laboratory operations.
Key points are:
- Suppliers should be selected and formally
evaluated to ensure that services and supplies have adequate
quality.
- Records of the selection and evaluation
process should be maintained.
- The quality of incoming material should
be verified against predefined specifications.
Service to the Customer
This chapter should ensure that the laboratory
continually meets customer requirements.
Key points are:
- The laboratory should cooperate with
customers to clarify their requests and to listen to their
inputs.
- The laboratory should have a formal
program to get ongoing feedback from customers.
- The laboratory should allow customers to
audit the laboratory.
Complaints
This chapter should ensure that any customer
complaints are documented, evaluated and adequately followed up.
Key points are:
- There should be a policy and procedure
for the resolution of complaints received from customers.
- Records of complaints and all steps taken
when resolving the complaint should be maintained. This includes
documentation of investigations and corrective actions.
Control of Nonconforming Testing and/or
Calibration Work
Tests and calibrations and other laboratory
operations should conform with previously defined specifications
such as laboratory specifications or specifications as defined by
clients. This chapter should ensure that nonconforming test and
calibration results are adequately followed up and that corrections
are initiated.
Key points are:
- There should be a policy and process that
come into effect when results do not conform with procedures.
- Corrective actions should be taken
immediately to avoid reoccurrence.
- The significance of nonconforming work
should be evaluated, for example, the possible impact on other
testing or calibration work.
- If necessary customers should be
notified.
Improvement
This chapter should ensure that the
effectiveness of the management system is continually improved.
Key points are:
- Suggestions for improvements should be
taken from audit reports, analysis of data, customer complaints
and suggestions, corrective and preventive actions and
management reviews.
- Suggestions should be collected over time
and reviewed by management for suitable actions.
Corrective Action
This chapter should ensure that the root cause
of nonconforming work or deviations from laboratory and management
procedures are identified and that adequate corrective actions are
selected, implemented, documented and monitored.
Key points are:
- Corrective actions can be triggered
through nonconforming tests or other work, customer complaints,
internal or external audits, management reviews and through
observations from staff.
- Corrective actions should be selected and
implemented to eliminate the specific problem and prevent
reoccurrence of the same problem.
- As the first step in the process the root
cause of the nonconformity should be identified.
- The effectiveness of the corrective
action should be monitored and evaluated.
Preventive Action
Preventive actions should be initiated when
potential sources of nonconformities have been identified.
Nonconformities could be technical or related to the management
system. The objective is to prevent reoccurrence of the same or
similar nonconformities.
Key points are:
- There should be a procedure to identify
potential sources of nonconformities and to define preventive
actions to prevent reoccurrence of these nonconformities.
- Preventive actions should not be limited
to one specific nonconformity, for example, due to the use of a
specific non-calibrated instrument for testing. The preventive
action plan should ensure that non-calibrated instruments will
not be used for testing.
- The effectiveness of the preventive
action should be monitored and evaluated.
Control of Records
This chapter should ensure that all records in
a laboratory are uniquely identified, that they are readily
available when needed and that they are protected against
non-authorized access for viewing or changing.
Key points are:
- There should be procedures for
identification, collection, indexing, storage, retrieval and
disposal of records.
- Records should be stored such that their
security, confidentiality, quality and integrity are ensured
throughout the required retention time.
- For technical records, e.g., test reports
of analytical measurements, original observations should be
retained together with processing parameters that would allow
tracking final results back to the original observations.
- Record format can be hard copies or
electronic media. There should be procedures to protect and
back-up electronic records and to prevent unauthorized access.
- Records can be corrected if there are
mistakes. In this case the original record should be crossed out
and still be visible.
- When electronic record systems are used,
the same principle applies. The laboratory should ensure that
original records are not overwritten by the system and that
corrections are recorded together with the original records. It
is recommended to use a system that prevents overwriting
original records and stores changes in an electronic audit trail
that can be viewed and printed.
Internal Audits
Internal audits should verify that the
laboratory complies with ISO/IEC 17025 and with internal technical
and quality procedures. Internal audits are also an excellent
preparation for external assessments and can help to continually
improve the quality system.
Key points are:
- The laboratory should have a procedure
and a schedule for internal audits. Internal audits can either
cover the whole laboratory and all elements of the quality
system at one specific period of time or can be divided into
several subsections.
- The schedule should be such that each
element of the quality system and each section of the laboratory
are audited yearly.
- The audit program should be managed by
the quality manager.
- Audit findings related to the quality of
test and calibration results should be reported to customers.
- •Audit follow-up activities should
include corrective and preventive action plans (CAPA). The
effectiveness of the plans should be monitored.
Management Reviews
Requirements in this chapter should ensure the
continued suitability and effectiveness of the quality system,
policies and testing and calibration procedures.
Key points are:
- here should be a schedule and procedure
for periodic management reviews.
- Recommended review frequency is once a
year.
- he management review should include a
discussion on the outcome from recent internal audits and
external assessments, corrective and preventive actions, results
of proficiency testing, customer complaints and feedback and any
recommendations for improvements.
- Management should decide on follow-up
activities. Such activities should be monitored for
effectiveness.
Technical requirements address the competence
of staff, sampling and testing methodology, equipment and quality
and reporting of test and calibration results. This chapter is
divided into ten sections.
General
The technical requirements part starts with a
general chapter. It makes readers aware of the fact that the
correctness and reliability of test and calibration results are
determined by a variety of factors.
Key points are:
- The different factors impacting the
quality of results should be documented. They include, for
example, sampling, equipment, test methods and environmental
conditions.
- The extent to which the factors can
contribute to the measurement uncertainty should be used when
developing test and calibration methods.
Personnel
Personnel probably have the highest impact on
the quality of test and calibration results. This chapter should
ensure that all laboratory personnel who can impact test and
calibration results are adequately qualified for their job.
Key points are:
- Only competent personnel should perform
testing and calibrations. This includes part-time as well as
full-time employees and all management levels.
- • Competence can come from education,
experience or training.
- Management should define and maintain
tasks, job descriptions and required skills for each job.
- Based on required skills and available
qualifications a training program should be developed and
implemented for each employee.
- The effectiveness of the training should
be evaluated. If the training is related to a specific test
method, the trainee can demonstrate adequate qualification
through successfully running a quality control or proficiency
test sample. A statement from the trainee such as ‘I have read
through the test procedure’ is not enough.
- Management should authorize personnel to
perform specific tasks, for example, to operate specific types
of instruments, to issue test reports, to interpret specific
test results and to train or supervise other personnel.
- The date of this authorization should be
recorded. The associated tasks are not performed before the
authorization date.
Accommodation and Environmental Conditions
This chapter has been included to ensure that
the calibration and test area environment will not adversely affect
the measurement accuracy. It includes five sections which are mostly
common sense. For example, one clause recommends having effective
separation between neighboring areas when the activities therein are
incompatible. An example would be to separate laboratories which
analyze extremely low traces of a solvent from those which consume
large quantities of the same solvent for liquid-liquid extraction.
Key points are:
- Environmental conditions should not
adversely affect the required quality of tests. This, for
example, means that equipment should operate within the
manufacturer’s specifications for humidity and temperature.
- The laboratory should monitor, control
and record environmental conditions. Special attention should be
paid to biologic sterility, dust, electromagnetic disturbances,
radiation, humidity, electrical supply, temperature, sound and
vibration. Tests should be stopped when the environmental
conditions are outside specified ranges.
- Areas with incompatible activities should
be separated.
- Access to test and calibration areas
should be limited to authorized people. This can be achieved
through pass cards.
Test and Calibration Methods and Method
Validation
Accurate test and calibration results can only
be obtained with appropriate methods that are validated for the
intended use. This chapter deals with the selection and validation
of laboratory-developed and standard methods and measurement
uncertainty and control of data.
Key points are:
- Methods and procedures should be used
within their scope. This means the scope should be clearly
defined.
- The laboratory should have up-to-date
instructions on the use of methods and equipment.
- If standard methods are available for a
specific sample test, their latest edition should be used.
- Deviations from standard methods or from
otherwise agreed methods should be reported to and agreed by the
customer.
- When using standard methods, the
laboratory should verify its competence to successfully run the
standard method. This can be achieved through repeating one or
two critical validation experiments and/or through running
method specific quality control and/or proficiency test samples.
- Standard methods should also be validated
if they are partly or fully out of the scope of the test
requirement.
- Methods as published in literature or
developed by the laboratory can be used, but should be fully
validated. Clients should be informed and agree to the selected
method.
- Introduction of laboratory developed
methods should be introduced following a plan.
- The following parameters should be
considered for validating in-house developed methods: limit of
detection, limit of quantitation, accuracy, selectivity,
linearity, repeatability and/or reproducibility, robustness and
linearity.
- Exact validation experiments should be
relevant to sample and required information.
- Sometimes, standard and in-house
developed validated methods need to be adjusted or changed to
ensure continuing performance. For example, the pH of a HPLC
mobile phase may have to be changed to achieve the required
separation of chromatographic peaks. In this case the influence
of such changes should be documented, and if appropriate, a new
validation should be carried out.
- Validation includes specification of the
requirements and scope, determination of the characteristics of
the methods, appropriate tests to prove that the requirements
can be fulfilled by using the method and a statement on the
validity.
Key points for measurement uncertainty are:
- The laboratory should have a procedure to
estimate the uncertainty of measurement for calibrations and
testing.
- For uncertainty estimation the laboratory
should identify all the components of uncertainty.
- Sources contributing to the uncertainty
can include the reference materials used, the methods and
equipment used for sampling and testing, environmental
conditions and personnel.
Key points for control of data are:
- Calculations used for data evaluation
should be checked. Checking calculations is best done during
software and computer system validation. As an example,
spreadsheet formula as defined by a specific user should be
verified with an independent device, for example, a handheld
calculator. Data transfer accuracy should be checked. Accuracy
of data transfer between computers can be automatically checked
with MD5 hash sums.
- Computer software used for instrument
control, data acquisition, processing, reporting, data transfer,
archiving and retrieval developed by or for a specific user
should be validated. The suitability of the complete computer
system for the intended use should be validated.
- Any modification or configuration of a
commercial computer system should be validated. Examples for
such configurations are defining report layouts, setting up IP
addresses of network devices and selecting parameters from a
drop down menu.
- Electronic data should be protected to
ensure integrity and confidentiality of availability of
electronic records. For example, computers and electronic media
should be maintained under environmental and operating
conditions to ensure integrity of data.
Equipment
Well-functioning, performing and maintained
equipment are prerequisites to ensure ongoing accurate test and
calibration results. This chapter deals with the capacity and
quality of equipment. The whole idea is to make sure that the
instrument is suitable to perform selected tests/calibrations and is
well characterized, calibrated and maintained.
Key points are:
- Equipment should conform to
specifications relevant to the tests. This means that equipment
specifications should first be defined so that when conforming
to defined specifications the equipment is suitable to perform
the tests.
- Equipment and its software should be
identified and documented.
- Equipment should be calibrated and/or
checked to establish that it meets the laboratory's
specification requirements.
- Records of equipment and its software
should be maintained and updated if necessary. This includes,
for example, version numbers of firmware and software. It also
includes calibration and test protocols.
- The calibration status should be
indicated on the instrument together with the last and the next
calibration.
Measurement Traceability
Traceability of equipment to the same standard
is a prerequisite for comparability of test and calibration results.
Ideally all measurements should be traceable to SI units. While this
is typically possible for physical measurements such as length (m)
and weight (kg) this is more difficult in chemical measurements.
Key points for traceability of calibrations
are:
- Calibration of equipment should be
traceable to the International System of Units (SI).
- Traceability of laboratory standards to
SI may be achieved through an unbroken link of calibration
comparisons between the laboratory standard, secondary standard
and primary or national standard.
- If traceability to SI units is not
possible, the laboratory should use other appropriate
traceability standards. These include the use of certified
reference material and the use of consensus standards or
methods.
Sampling
This chapter has been added to ensure that
statistically relevant representative samples are taken and that all
information on the sample and the sampling procedure is recorded and
documented.
Key points for sampling are:
- Sampling should follow a documented
sampling plan and sampling procedure.
- The sampling plan should be based on
statistical methods.
- The sampling procedure should describe
the selection and withdrawal of representative samples.
- The sampling location and procedure, the
person who took the sample and any other relevant information
about the sampling location should be recorded.
Handling Test and Calibration Items
This chapter should ensure that the sample
integrity is maintained during transport, storage and retention and
that it is disposed of safely.
Key points for handling test and calibration
items are:
- Test and calibration items should be
uniquely identified.
- Sample transportation, receipt, handling,
protection, storage, retention and/or disposal should follow
documented procedures.
- The procedures should prevent sample
deterioration and cross-contamination during storage and
transport.
Assuring the Quality of Test and Calibration
Results
This chapter should ensure the quality of
results on an ongoing basis, for example, through regular analysis
of quality control samples.
Key points are:
- The validity of test results should be
monitored on an ongoing basis.
- The type and frequency of tests should be
planned, justified, documented and reviewed.
- Quality control checks can include the
regular use of certified reference materials, replicate tests or
calibrations using the same or different methods and retesting
or recalibration of retained items.
Reporting of Results
This paragraph describes how test/calibration
results should be reported. This is important for an easy comparison
of tests performed in different laboratories. The chapter has some
general requirements on test reports such as clarity and accuracy
but it also has very detailed requirements on the contents.
Test reports and calibration certificates
should include:
- The name and address of the laboratory.
- Unique identification of the test report
or calibration certificate (such as the serial number).
- The name and address of the client.
- Identification of the method.
- A description and identification of the
item(s) tested or calibrated.
- Reference to the sampling plan and
procedures used by the laboratory.
- The test or calibration results with the
units of measurement.
- The name(s), function(s) and signature(s)
or equivalent identification of person(s) authorizing the test
report or calibration certificate.
- Statement on estimated uncertainty of
measurement (for test reports (‘where applicable’).
- When opinions and interpretations are
included, the laboratory shall document the basis upon which the
opinions and interpretations have been made.
- Opinions and interpretations shall be
clearly marked as such in a test report or calibration
certificate.
Now that you have an overview on required management and technical
controls we will give recommendations on how to efficiently comply
with some key requirements. The scope of this tutorial does not
allow going into all the details and we cannot cover all
requirements. We will focus on topics which most likely are new to
laboratories without a quality system. These include: specific
organizational structure, formal equipment calibration and testing,
measurement traceability and uncertainty. We try to balance missing
detailed information by providing references to official guidelines,
textbooks and other literature. For example, EURACHEM/CITAC, EUROLAB
and ILAC have developed guidance documents for measurement
uncertainty and traceability in measurement (3-9). ISO has published
a "Guide to the Expression of Uncertainty in Measurement" (10). A
EURACHEM-UK working group has developed a guidance for qualification
of analytical instruments which was published in "Accreditation and
Quality Assurance" (11). EUROLAB has published a technical report
called "Management of Computers and Software in Laboratories with
Reference to ISO/IEC 17025:2005" (12). Huber has authored a text
book “Validation and qualification in Analytical Laboratories” and
Thompson et al. gave recommendations for an "International
Harmonized Protocol for Proficiency Testing of Chemical Analytical
Laboratories" (14). Help is also readily available from
accreditation bodies. For example, A2LA has a "Policy on Measurement
Traceability" (15) and Labcompliance provides a complete "ISO/IEC
Accreditation Package" with an example master plan, SOPs, forms and
checklists (2).
Organizational Structure
ISO/IEC requires that organizational arrangements should be such
that departments with conflicting interests do not adversely
influence compliance with the standard. For example, finance and QA
should operate independently from laboratory activities. Figure 2
shows an example of an organization chart. Finance and QA do not
report to laboratory management but rather to the director of the
company.

Figure 2: Example for Organizational Structure
(from Reference 2)
Equipment
Each laboratory should have a plan on how to ensure adequate
equipment functioning and performance before and during sample
measurement. Main activities are calibration, checks to verify
specified performance and maintenance.
The extent of testing depends on the
complexity and use of the equipment. Each laboratory should have
processes on how calibration and testing is performed for different
types of equipment. Preferably the whole range of equipment should
be divided into a few categories, for example, A, B and C and
calibration and/or verification tests should be associated with each
category. Instruments in the simplest category A such as stirrers
and mixers usually don’t need to be tested but only visually
inspected. Category B instruments such as balances and pH meters
should be calibrated according to manufacturer SOPs and more complex
instruments such as gas chromatographs should be fully tested
according to their intended use.
In the following paragraphs we give
recommendations on how to specify, test and maintain analytical
equipment for ISO/IEC 17025 compliance.
Documenting Specifications
ISO/IEC 17025 requires that equipment and
software should comply with specifications that are relevant to the
tests. Therefore, the first step in the process is to define and
document the equipment specifications.
- For simple equipment such as balances and
pH meters the use of manufacturer specifications is recommended.
- For more complex equipment hardware such
as gas chromatographs or mass spectrometers the manufacturer’s
specifications can also be used. This is only recommended as
long as all vendor-specified functions are required by the
intended applications over the full specified range.
Alternatively the user can define his/her own specification
according to the intended use of the instrument.
- Commercial software and computer systems
typically provide more functionality than required by a specific
user. Therefore, for computer systems, the user should define
his/her own specifications according to the system use. A
functional specifications list will help define user
specifications.
Selecting a Vendor
Selecting equipment suppliers should follow a
documented procedure and well-defined criteria. Such criteria are:
- Vendor’s equipment meets the user’s
requirement specifications
- Leading position of the vendor in the
marketplace.
- Design, development and manufacturing of
equipment and software in a quality system environment, e.g.,
ISO 9001.
- Installation, familiarization and
training services
- Metrology based calibration and
functional testing services through qualified engineers
- Phone and onsite support in local
language.
Installation and Documentation
Installation can be performed by the vendor or
by the user. Steps include:
- Verification that the location meets the
environmental specifications as defined by the vendor.
- Installation of equipment hardware
according to vendor specifications.
- Installation of software and start-up
according to vendor specifications
- Documentation of hardware and software,
e.g., vendor, product number, model number, serial number and
location.
Initial Testing for Calibration
and/or Performance Verification
Equipment used for measurement should be
tested before initial use to ensure its suitable performance. This
is performed through calibration, for example, the mass of a
balance, or through verification with specified performance
characteristics, for example, sampling precision of a gas
chromatograph.
Steps for testing include:
- Develop test procedures and test
protocols.
- Define acceptance criteria based on
documented specifications
- Select and order traceable test tools,
for example, reference weights for calibration of a balance.
- Make sure that test engineers have the
right qualifications to perform the tests.
- Perform the tests and document test
results.
- Verify that acceptance criteria are met.
- Label the equipment with the status,
e.g., last and next calibration.
- Maintain records of calibration and
checks.
These tests can be performed by the vendor or
by the user. Advantages of testing by the vendor can be demonstrated
using the Agilent Functional Verification Service (FVS) for gas and
high-performance liquid chromatography as example . This service was
specifically developed to meet ISO/IEC 17025 requirements.
- Agilent has decades of knowledge and
experience with factory and field testing of equipment. As a
result the selection and sequence of the test is optimized for
highest speed and lowest instrument downtime, without comprising
accuracy and calibration specifications.
- Agilent engineers bring along calibrated
test tools that meet traceability requirements.
- Agilent engineers also bring certificates
to document qualification.
- Agilent’s tests are combined with
recommended preventive maintenance for all the critical
functional components of the equipment.
- Agilent provides global lab-to-lab
consistency via standardized and well recognized maintenance and
verification testing protocols.
- Agilent’s calibration and test
certificates are globally well recognized by internal auditors
and official assessors.
Testing During Ongoing Use
Equipment characteristics and performance can
change over time. Therefore, the equipment quality program should
have ongoing measures to prove that the equipment is suitable for
its intended use, which means equipment should be routinely tested.
The type and frequency of such tests depends
on the equipment. For example, a balance is checked daily with
laboratory reference weights and a chromatograph can be checked
using well-defined quality control samples.
In addition to the more frequent tests that
only challenge a subset of all specifications, it is recommended to
repeat all the initial tests as described above, for example, for
chromatographs, once a year. A balance is also typically calibrated
once a year by the vendor with calibrated and traceable standard
weights.
Maintenance and Repair
The laboratory should have a maintenance plan
to carry out preventive maintenance activities and a procedure for
unplanned repair to ensure ongoing performance and reliability.
- Defective equipment should not be used
for tests and calibration. Smaller devices should be taken out
of the laboratory and bigger instruments should be clearly
labeled as being defective. Specified functioning and
performance should be verified after repair.
- Records of maintenance and repairs should
be maintained.
Software and Computer Systems
ISO/IEC 17025 requires computer systems and
software used for acquisition, processing, recording, reporting,
storage and retrieval of test and calibration data to be validated
when the software is developed, configured or customized by the
user. Commercial Off-the-Shelf (COTS) software, for example, word
processors or databases may be considered to be validated.
EUROLAB has developed a technical report with
recommendations for "Management of Computers and Software in
Laboratories with Reference to ISO/IEC 17025:2005" (12). The report
not only advises on validation steps for different software and
system risk categories but also has recommendations on how to ensure
security, availability, integrity and confidentiality of electronic
records.
The report divides software into five
categories:
- Operating systems.
- Firmware as built into automated
equipment.
- Standard software packages, e.g., word
processors and non-configurable computerized analytical systems.
- Configured software packages, e.g., Excel
formulae and configurable computerized analytical systems.
- Custom built software.

Figure 3 shows simplified validation
activities for all five categories.
Measurement Traceability
ISO/IEC 17025 requires reference material used for calibration of
measurement equipment to be traceable to SI Units, where possible.
Typically laboratories use their own internal reference material for
calibration. Traceability of such material to SI units can be
achieved through an unbroken chain of comparisons between laboratory
reference material and SI units. An example is shown in Figure 4.

Figure 4: Traceability Chain of Laboratory
Reference Material
Working standards are regularly compared by an accredited laboratory
with secondary standards which are calibrated by a national
metrology institute or an accredited reference laboratory. For this
kind of comparison the measurement uncertainty should be known and
documented in calibration certificates so that the measurement
uncertainty of the working standard can be estimated and reported.
This concept works well for physical
measurement such as meter (m) for length, kilogram (kg) for mass and
Kelvin (K) for temperature. For this reason reference balance
masses, thermometers and thermocouples should be traceable to SI
Units through an unbroken chain of comparisons performed by
accredited laboratories and national metrology institutes.
For most reference material used for chemical
measurements traceability to SI units is very difficult and not
practical. The traceability chain in Figure 4 ends at the lower
level, e.g., at suppliers of standard reference material, for
example, NIST, at suppliers of certified reference material or at a
company’s accredited metrology laboratory. When traceability to SI
is not possible, ISO/IEC 17025 recommends the use of certified
reference material that is provided by a competent supplier and is
well characterized. Alternatively well-defined methods also called
primary or definitive methods as agreed by all parties can be used
to establish traceability. This topic has been discussed in
different working groups. For example, detailed recommendations have
been published by ILAC (7) and EURACHEM/CITAC (8). X.R. Pan (17)
suggested a classification scheme of reference material used for
chemical measurements. The classification as described below is well
accepted in chemical laboratories.
Primary Reference Material
- Also called primary standards.
- Developed by a national metrology
laboratory.
- Certified by primary/consensus method.
- Traced back to SI units and/or verified
by international comparison
Certified Reference Material
- Also called secondary standards.
- Derived from primary reference material
with statement of uncertainty.
- Usually prepared by a specialized
reference laboratory.
- Certified by reference methods or
comparison methods.
- Recognized by national or otherwise
specialized authoritative organization.
- Working Reference Material
- Also called internal reference material.
- Derived from certified reference
material.
- Accuracy verified by well characterized
and validated methods.

Measurement Uncertainty
There is an uncertainty associated with every
test and calibration. For testing, this occurs from errors arising
at the various stages of sampling, sample preparation, measurement
and data evaluation. In other words, whenever any quantitative
measurement is performed, the value obtained is only an
approximation of the true value. Users of the measurement data
should have an idea of how much the reported result may deviate from
the true value.
ISO/IEC 17025 recommends the results of
quantitative measurement to be reported as both a single value and
together with the possible deviation from the true value. This is
logical for any report with quantitative results. It is, for
example, of no use if a report on a food sample states 0.1 percent
of compound X, and the user of the data is still unsure whether this
could be 0.05 or 0.4 percent. An uncertainty statement provides the
user with information on the approximate measurement tolerances and
the expected limits within which the true value of the measurement,
such as analyte concentration, is supposed to lie. Without such
documentation, although the analyst can estimate the level of
uncertainty many times, the client or user of the data cannot.
Information on uncertainty is of particular
importance if a specification limit is to be verified and reported.
For example, if according to a purchasing agreement, a product can
only be released if compound X is below 0.5 percent, the test report
may not contain a statement about compliance if the measurement
results extended by the measurement uncertainty is above 0.5
percent. When parameter(s) are claimed to
be within a specified tolerance the measurement value(s) extended by
the estimated uncertainty of measurement shall fall within the
specification limit.
ISO has published a "Guide to the Expression
of Uncertainty in Measurement" (10). It establishes general rules
for evaluating and expressing uncertainty in measurement across a
broad spectrum of measurements. EURACHEM has produced an excellent
document containing much more detail on how the concepts of the ISO
guide can be applied in chemical measurement (4). The whole process
of measurement uncertainty is schematically shown in Figure 5. The
basic ideas are explained in this tutorial, but for more detailed
information, readers are encouraged to study the EURACHEM document
(4).
The concept of evaluating uncertainty is fairly straightforward. It
requires a detailed knowledge of the nature of the measurand and of
the measurement method, rather than an in-depth understanding of
statistics.

Figure 5: Estimating Measurement Uncertainty
- Develop the specifications by writing a
clear statement of exactly what is to be measured and the
relationship between this and the parameters on which it
depends. For example, if the measurement temperature has an
influence on the result, the measurement temperature should also
be defined.
- Develop a workflow diagram for the entire
sampling, sample preparation, calibration, measurement, data
evaluation and data transcription process (see Figure 1 for
analytical sample testing).
- Identify and list sources of uncertainty
for each part of the process or for each parameter. Possible
sources for errors may be derived from non-representative
sampling, operator bias, a wrongly calibrated instrument, lack
of ideal measurement conditions, chemicals with impurities and
errors in data evaluation.
- Estimate and document the size of each
uncertainty, for example, as standard deviations or as RSDs.
These data should be gathered from a series of measurements.
Where experimental evaluation is impossible or impractical, the
individual contributions should be estimated from whatever
sources are available. Sources for this kind of estimation can
be found in the supplier’s information or in the results of
interlaboratory studies or proficiency testing. The procedures
and thoughts behind the way the contributions have been measured
or estimated should be documented.
- Combine separate contributions in order
to give an overall value. For example, where individual sources
of uncertainty are independent, the overall uncertainty can be
calculated as a multiple of the sum of squared contributing
uncertainty components, all expressed as standard deviations.
Computer software or spreadsheet programs can help to automate
this calculation.
The whole procedure should be documented in
such a way that sufficient information is available to allow the
result to be reevaluated if new information or data become
available. A complete documentation should include:
- A description of the methods used to
calculate the measurement result and its uncertainty from the
experimental measurements.
- The values and sources of all
corrections.
- A list of all components of uncertainty
with full documentation on how each of these was evaluated.
Reference 4 includes many practical examples
with data from different analyses, as well as formulas for
evaluating, calculating and reporting standard and expanded
uncertainty. Sample analysis reports should include an uncertainty
number, which is typically expressed as:
Result = x ± u (units)
or
Result = x (units)
Uncertainty = u (units)
ISO/IEC 17025 accreditation should be well
thought out and well prepared. It can be quite expensive but can
also have big benefits. The balance between costs and benefits
should be worked out and documented.
Going for ISO/IEC 17025 will impact the entire
laboratory and supporting services such as human resources,
documentation and finance departments. Therefore, while the decision
to initiate and fund the project will be made by management, all
affected departments should be involved in the process.
The entire process is divided into two phases:
Investigation phase and implementation phase. In the investigation
phase information is collected to decide if going for accreditation
makes business sense. Once the decision is made the laboratory
develops and implements documentation in preparation for the
accreditation assessment.
Figure 6 illustrates the steps for both phases.

Figure 5. Steps towards ISO/IEC 17025 accreditation
Investigation
- Management initiates, funds and otherwise
supports the investigation.
- Management nominates a project owner.
Ideally the person should be well experienced in laboratory
operations, have a good business understanding, good
communication skills and a good understanding of quality
systems.
- The project manager with support from
management recruits a project team. Members should come from the
laboratory management, QA, finance, human resources, training
and documentation groups.
- The project team defines the scope of the
intended accreditation. This could include all calibrations
and/or tests performed in a lab or just part of them.
- The project team studies the
accreditation requirements in detail. The main source is the
standard ISO/IEC 17025, that is supported by official
guidelines, other literature and by external expert advice.
- The project team develops a requirements
list. The list should also include all documents as required by
the standard, for example, policies, a quality plan and
procedures for most of the requirements.
- The project team prepares a gap analysis
through a comparison between ISO requirements as listed above
and what is already available and implemented. A gap exists
where existing policies, processes or procedures do not fully
meet the stated requirements. This should include all processes
and procedures for management controls and technical controls
such as for sampling, method validation, equipment calibration,
qualification and maintenance, people qualification and others.
- Using the outcome of the gap analysis the
project team develops a task list. The list is completed with
additional tasks such as selecting and dealing with an
accreditation body.
- The project team together with the help
of an external consultant makes an estimation of the overall
ISO/IEC implementation costs. This should not only include costs
for initial set-up but also for maintaining the quality system.
The costs are compared with direct and tangible estimated
additional return that comes from getting accreditation status.
Tangible returns are, for example, savings through more
efficient operation.
- The team makes a rough estimation of the
return of investment for both short and long term and makes a
recommendation to management.
- Management decides to accept or reject
the proposal and go for the accreditation or not.
Implementation
- If the decision is made to pursue
accreditation the project owner forms implementation teams for
different areas. It is most important that all affected
departments at all management levels are represented in the
teams.
- The project owner with the help of QA
searches for an accreditation body and selects the one that best
fits the laboratory´s needs. There are several possibilities to
find accreditation bodies. Probably the best way is to ask
accredited laboratories about their experiences.
- The teams develop documentation, e.g.,
procedures under the supervision of the project owner.
- The project owner arranges for staff
training.
- Quality assurance performs an internal
audit and initiates corrective actions, if necessary.
- The selected accreditation company
performs a pre-assessment.
- The project owner initiates corrective
actions.
- The accreditation company performs an
accreditation audit.
ISO/IEC requires different types of
documentation as illustrated in the documentation pyramid in Figure
7.

Figure 7: Documentation Pyramid
A policy documents the laboratory´s intent to
implement ISO/IEC 17025. The Quality Manual is the top tier of the
document hierarchy. It describes the approaches to achieve quality
data. It also includes policy statements describing the intent and
goal of the laboratory to conform with ISO/IEC requirements. For
example, a policy statement could be: All personnel involved in
calibration and testing should be competent for the assigned task.
A process or generic procedure describes how
various quality requirements can be achieved. For example, it
describes how the requirement ‘Personnel should be competent for the
assigned task’ can be implemented
Standard Operating Procedures or Working
Procedures are step-by-step instructions on how to exactly perform a
specific task, for example, how to calibrate a specific instrument.
Records are generated on a day-by-day basis. Examples are analytical
results from product tests or calibration records of a balance.
All documents should be well controlled, for
example, each change should be authorized and logged and the updated
document should get a new revision number or code.
Policies and Quality Manual
Policies including the quality policy
statement should be documented in the quality manual. The quality
policy statement should be written by senior management. It should
outline the laboratory’s commitment to quality. The quality manual
describes the quality system and documents the laboratory's goal and
overall concept on how to conform with ISO/IEC17025. It should also
describe how the remainder of the quality system documentation is
organized. It should be developed by working groups representing
different departments.
Processes
Processes or standard procedures describe how
various ISO/IEC 17025 requirements can be achieved. For example, it
describes how the requirement ‘All personnel involved in calibration
and testing should be competent for the assigned task’ can be
implemented. Another example is the laboratory´s approach to
calibrate and check different types of equipment. For a better
understanding, process flow charts should be included in a process
description.
(Standard) Operating Procedures and Work
Instructions
Routine activities follow documented
procedures. These are typically defined as standard operating
procedures and/or work instructions. While quality manuals and
processes describe the tasks and approaches, procedures and work
instructions give step-by-step instructions on how to do the tasks.
Examples for SOPs are procedures for checking and calibration of
equipment. All laboratory SOPs should use the same format, which
makes it easy for writing and reading. A good practice is to have a
SOP on how to author, review, approve, distribute and update SOPs.
Preferably SOPs should be written by senior members of anticipated
user groups. This helps ensure that SOPs have the right level of
information and are used and followed.
Records
Records to demonstrate conformity with ISO/IEC
17025 and as required by customers should be retained for a specific
amount of time. Examples are original laboratory observations, test
results, supporting documents such as chromatograms and also
training certificates and equipment calibration protocols.
Checklists, forms, templates and examples help
implement quality work effectively and consistently. Examples are
checklists and worksheets for vendor assessment, handling
nonconforming test results and for internal audits. They help
document specific tasks consistently and effectively.
Document Control
Development and maintenance of documentation
should be controlled through document control and management
procedures that are part of the management system. Documents include
internal and external documents. Examples for internal documents are
SOPs, quality manuals and training plans. Examples for external
documents are regulations, standards, test methods and instrument
operating manuals.
The procedure for document control should
ensure that:
- Official documents are created or
acquired, reviewed and approved prior to use.
- Documents are uniquely identified with
document and revision number, date of revision and issuing
authority.
- A quality list with all controlled
documents is maintained by QA. The list includes document and
revision number, title, date of issue, date of last review and
locations.
- Internal documents include page number
and total number of pages on each page.
- Users of the documents are adequately
trained before the documents are released for use.
- Current authorized versions are readily
available at the user’s workspace.
- Documents are reviewed according to a
schedule and revised to ensure suitability and ongoing
conformance with regulations and internal procedures.
- Invalid and obsolete documents are
promptly removed from all points of issue or use, or marked as
uncontrolled to ensure that only current authorized versions of
appropriate documents are available for active use at relevant
locations.
- Obsolete documents retained for either
legal or knowledge preservation are marked as ‘Archived’, dated
and signed. The retention period for the documents conforms with
internal procedures.
- Changes to a document are reviewed,
approved and communicated to users.
- The changes are recorded in a document
change log. The log information indicates the reason and the
nature of the change.
- When documents are created, signed and
maintained in electronic form, the computer system and records
comply with national or international regulations and guidelines
for electronic or digital signatures.
Internal audits are managed by the quality
manager. They verify conformance with the ISO/IEC 17025 requirements
and with company policies, processes and procedures. They are also
quite useful in preparation for external audits. External auditors
can come from clients or from accreditation bodies. They verify that
the laboratory is operating in compliance with ISO/IEC 17025.
There should be procedures that clearly
outline who is doing what before, during and after internal and
external audits. Overall owners should be defined and all employees
who may be affected by the audit should be trained. This chapter
summarizes recommendations for audits. To make best use of internal
audits they should be designed, executed and followed up very much
in the same way as external audits are expected to be. The
recommendations for preparation, conduct, documentation and
follow-up are written for the audited departments, not for the
auditors.
Internal Audit Schedule
Internal auditing should follow a
predetermined schedule covering all activities over a reasonable
period of time. It is inconvenient to audit all activities in a
single audit so it should be spread over several quarterly or
monthly audits. The schedule for such audits is conveniently drawn
as a matrix covering, for example, a year in which dates are set for
each part of the quality system. Audit schedules can be organized as
horizontal or vertical. A vertical audit checks compliance of, for
example, a single test through all steps and records from sampling
to archiving of records. A horizontal audit examines every aspect of
a single requirement, for example, equipment. shows an example for a
horizontal audit.

Figure 8: Example for Horizontal Audit
Schedule
Preparation
- Assign an overall owner and host for the
audit.
- Assign a technical contact to get access
and review the completeness of records and other documentation
for items to be audited. The assigned technical contact should
be present all the time.
- Set up a work area for the inspectors.
- Review the schedule.
- Prepare and train staff.An audit may be a
tough experience for all people involved. They therefore need to
be informed on what will happen and what questions may be asked.
Conduct
- Maintain a continuous log of the audit.
- Provide copies (do not give originals
away).
- Keep duplicates of all information
supplied to auditors.
- Take immediate corrective action, when
appropriate.
- Hold a daily debriefing meeting to assess
the progress.
- Keep all documents in the work area.
- Accompany the auditor all the time.
- Be courteous and cooperative.
- Answer only questions that are asked.
- If you are unable to answer, tell the
auditor openly.
- Protect proprietary information.
Close
- Clarify any open questions or cause for
dissatisfaction in the exit meeting.
Follow-up
- Develop a corrective action plan (owners,
tasks, deliverables and schedule).
- Develop a preventive action plan (owners,
tasks, deliverables and schedule).
- Monitor the plan.
Laboratories are frequently faced with a situation where they have
to comply with both regulations and quality standards at the same
time.
Examples are:
- A clinical laboratory performs contract
analyses from pre-clinical and clinical studies for
pharmaceutical companies. The laboratory also performs special
tests for hospitals. The laboratory has to operate in compliance
with US FDA and EU GLP and GCP regulations for clinical and
pre-clinical study tests. Some customers also require laboratory
accreditation according to ISO/IEC 17025, others also according
to ISO 15189, a standard for medical laboratories.
- A chemical company is certified for ISO
9001. The scope of the certification also covers the analytical
service laboratory. In addition, the laboratory performs
contract analyses for other companies and has received
laboratory accreditation in compliance with ISO/IEC 17025. The
laboratory has to work in compliance with ISO 9001 and with ISO
Guide 17025.
- An independent test laboratory performs
GLP studies as a subcontractor for a pharmaceutical company.
Occasionally, the laboratory also performs analyses for
pharmaceutical manufacturing control departments. The laboratory
has also received laboratory accreditation for specific food
analyses according to ISO/IEC 17025. The laboratory has to
comply with ISO/IEC 17025 and with GLP and cGMP regulations.
International companies frequently face this
kind of problem. Their laboratories not only have to comply with
regulations from different countries but also, simultaneously, with
quality and accreditation standards. The solution to this problem is
to combine all regulations and quality standards in a single quality
manual and a single set of operating procedures. The recommended
documents and how they relate to each other are shown in Figure 7.
The quality manual should place the company’s
own quality system first and foremost. This may be based on a
well-known laboratory quality standard, such as ISO/IEC 17025.
The quality manual and operating procedures
should include aspects of various regulations and quality standards
applied within the company. For specific requirements only required
by single regulations the quality manual and procedures should
include sections that only apply to those particular regulations.
For example, in the ‘responsibility’ section it should mention that
for GLP studies the function of a study director is required. The
tasks and responsibilities should be described in a SOP.
- ISO/IEC 17025, General Requirements for
the Competence of Testing and Calibration Laboratories, 2005.
- Labcompliance, ISO/IEC17025 Accreditation
Package, 2009.
http://www.labcompliance.com/books/iso17025
- EURACHEM/CITAC Guide, Use of Uncertainty
Information in Compliance Assessment, 2007.
- EURACHEM/CITAC Guide CG4, Quantifying
Uncertainty in Analytical Measurement, ISBN 0-948926-15-5, 2000.
- EUROLAB, Measurement Uncertainty
Revisited: Alternative Approaches to Uncertainty Evaluation,
2007.
- ILAC, Introducing the Concept of
Uncertainty in Association with the Application of the Standard
ISO/IEC 17025, 2002.
- ILAC Policy on Traceability of
Measurement Results, 2002.
- EURACHEM/CITAC Guide, Traceability in
Chemical Measurement: A Guide to Achieving Comparable Results in
Chemical Measurement, 2003.
- EUROLAB, Guide to the Evaluation of
Measurement Uncertainty for Quantitative Test Results, 2006.
- ISO Guide 98-3, Uncertainty of
measurement -- Part 3: Guide to the Expression of Uncertainty in
Measurement (GUM:1995), Geneve, Switzerland, 2008.
- P. Bedson and M. Sargent, The Development
and Application of Guidance on Equipment Qualification of
Analytical Instruments, Accreditation and Quality Assurance, 1
(6), 265-274 (1996).
- EUROLAB Technical Report 2/2006,
Management of Computers and Software in Laboratories with
Reference to ISO/IEC 17025:2005.
- L. Huber, Validation and Qualification in
Analytical Laboratories, Interpharm, Informa Healthcare, New
York, USA, 1998. Second revision 2007.
- M. Thompson, S. Ellison and R. Wood,
International Harmonized Protocol for Proficiency Testing of
Chemical Analytical Laboratories, Pure Appl. Chem., Vol. 78, No.
1, pp. 145–196, 2006,
www.iupac.org/publications/pac/2006/pdf/7801x0145.pdf.
- American Association for Laboratory
Accreditation, A2LA Policy on Measurement Traceability, 2005.
- ISO/IEC Guide 2, Standardization and
Related Activities: General Vocabulary, 2005.
- X. R. Pan, Hierarchy of Reference
Materials Certified for Chemical Composition, Metrologia 34
(1997) 35-39.
To be updated on an ongoing basis
This chapter will be updated based on
questions and contributions from visitors
Q: Is ISO 17025 Accreditation required and
sufficient to comply with FDA Regulations, such as GLP and GMP?
Most requirements are the same or very
similar, but ISO 17025 is neither required nor sufficient. However,
many FDA laboratories have ISO 17025 implemented.
Rev 2: 2005
Q: When has ISO/IEC 17025 been introduced ?
Rev 1: 1999
Rev 2: 2005
Q: Is there any website where you can view or
download the ISO 17025 Standard?
ISO Standards are copyright protected. I am
not aware of a website with free download.
Q: Which organizations can award ISO 17025?
There are many in each country. To find them
in your area, use Google Search.
Q: What to do in case there is no proficiency
testing scheme for my application?
Do everything you can to proof accuracy of
your analytical data. Mechanisms include: extensive use of certified
reference material, verification of accuracy with an independent
analysis method and cross check of results within other departments
in you organization. Document the steps you took.
Q: Is an ISO/IEC 17025 accredited laboratory
also automatically ISO 9001 Certified?
ISO 17025 has a statement that testing and
calibration laboratories that comply with the International Standard
will also operate in accordance with ISO 9001.
On the other hand it states that
demonstrated conformity to this International Standard does not
imply conformity of the quality management system within which the
laboratory operates to all the requirements of ISO 9001. This means,
ISO 17025 conforms to related requirements of ISO 9001, but it does
not contain all the requirements as specified in ISO 90001. So the
answer to question above is 'no'.
This chapter will be updated on an on-going
basis