ISO/IEC 17025 in Laboratories

1. Management Summary
ISO/IEC 17025 is the global quality standard for testing and
calibration laboratories. It is the basis for accreditation from an
accreditation body. The current release was published in 2005.
There are two main clauses in ISO/IEC 17025 –
Management Requirements and Technical Requirements. Management
requirements are related to the operation and effectiveness of the
quality management system within the laboratory, and this clause has
similar requirements to ISO 9001. Technical requirements address the
competence of staff; testing methodology; equipment and quality; and
reporting of test and calibration results.
Implementing ISO/IEC 17025 has benefits for
laboratories, but the work and costs involved should be considered
before proceeding..
Main Benefits of Correctly Implemented
ISO/IEC 17025:
Implementing ISO/IEC 17025 has benefits for
laboratories, but the work and costs involved should be considered
before proceeding.
Implementing ISO/IEC 17025 as part of
laboratory quality initiatives provides both laboratory and business
benefits such as:
- Having access to more contracts for
testing and/or calibration. Some public and private
organizations only give contracts to accredited laboratories.
Accreditation will also help to get more contracts from
organizations that don’t mandate accreditation, but do give
preference to accredited laboratories in competitive situations.
- Improved national and global reputation
and image of the laboratory.
- Continually improving data quality and
laboratory effectiveness.
- Having a basis for most other quality
systems related to laboratories, such as Good Manufacturing
Practices and Good Laboratory Practices.
Analytical testing laboratories seeking
ISO/IEC 17025 will be impacted in multiple areas. The main
difference between good analytical practices and formal
accreditation is the amount of documentation to be developed. There
is no doubt that any good analytical laboratory uses qualified
analysts, checks the performance of equipment used for testing, and
validates analytical methods. However, many times the outcome of the
tests is not fully documented. ISO/IEC 17025 accreditation requires
formal documented environment – ‘what is not documented is a rumor,’
and is viewed by assessors as ‘not being done.’
The overall impact of accreditation on an analytical laboratory can
be best illustrated by looking at the whole sample/data workflow.
Figure 1 shows a typical laboratory workflow of samples and test
data, together with ISO/IEC 17025 requirements.

Figure 1: ISO/IEC 17025 Requirements for
Testing Laboratories
Requirements Overview
Requirements along the analytical workflow
- Sampling should be performed according to
a sampling plan, and all sample details should be documented
- Samples should be uniquely identified and
the sample integrity should be protected during transport and
storage.
- The quality of test results should be
monitored.
- Test reports should include test results
as well as an estimation of the overall measurement uncertainty.
The report should also include either detailed information about
the sample and test conditions, or a link to a reference
document.
- Records should be properly maintained to
ensure data integrity and availability.
Some requirements impact more than one
workflow step:
- All analytical methods and procedures
should be validated. This includes methods and procedures for
sampling, testing and data evaluation.
- Equipment used for sampling and testing
should be calibrated, tested, and well maintained. Material such
as calibration standards should be qualified and traceable to
System International (SI) units or to certified reference
material.
- Nonconforming test results should be
documented and controlled.
- People should be qualified for their
assigned tasks through education, experience, or training.
- Environmental conditions such as
temperature, humidity, and electromagnetic interference should
be monitored and controlled.
- All routine tasks should be performed
according to written procedures.
Some additional requirements impact not only
sample analysis, but also the organization of the entire laboratory:
- Specific documents should be developed
and maintained, including individual policies and a quality
plan.
- Known existing problems should be
corrected and an action plan should be developed to avoid
recurrence of the same or similar problems.
- All complaints from clients should be
formally followed up.
- A formal program should be used to manage
suppliers, service providers, and subcontractors.
- The organizational structure should be
such that there are no conflicting interests that could impact
quality.
- Compliance with ISO/IEC 17025 and
internal procedures should be assessed during regular internal
audits.
Key Steps towards Accreditation:
There are eight key steps towards laboratory
accreditation:
- Management defines a project owner
- The project owner studies details of the
standard, supporting literature, and other relevant information
- The project owner defines the preliminary
scope of accreditation and works with laboratory professionals
to prepare a list with requirements
- The project owner and laboratory
professionals perform a gap analysis to determine the difference
between the requirements and what is currently implemented in
the laboratory.
- Based on the outcome of the gap analysis,
the project owner, laboratory
professionals, financing and documentation professionals, and
external consultants estimate the costs for accreditation
- Estimated costs are presented to
management, along with incremental opportunities.
- Management decides to proceed with
accreditation.
- The project owner leads implementation
steps.
The 17025 Accreditation Package: Minimal cost
compared to the return
Mr. Russell Hensel, President, Hensel
Environmental Biology Laboratories, USA, made it very clear: The
Labcompliance Accreditation Package gives a huge head start.
"The package did exceed my expectations. It gives us a HUGE head
start on the lab start up. The cost is minimal compared to the
return. EXCELLENT documentation. The package did exceed my
expectations". To learn more about the package and more
testimonials,
click here
Companies have to continuously deliver
high-quality products and/or services if they want to be successful
in the marketplace over the long term. Quality improvement has
become a key national and international business strategy. Most
companies are using quality systems as a method of assuring the
consistency and conformity of products or services to a defined set
of standards or customer expectations.
Quality Systems
Several quality system standards were
developed in various countries in the 1960s and 1970s. The
MIL-Q-9858A was established in the United States in 1963, and the BS
5750 was established in 1979 in the United Kingdom. These are
probably the two most important standards from this era. The ISO
9000 series of quality standards was established in 1987 for
implementing and maintaining a quality system. This standard is
internationally accepted and can be used as a criterion for
third-party quality assessment.
ISO/IEC 17025 – Laboratory Quality System
Laboratories play an important role in company
quality systems. The ISO/IEC 17025 (1) can be used as a standard to
develop and establish a quality system for a laboratory and also for
assessment by laboratory clients or third parties. The standard can
also be used as a criterion for laboratory accreditation. Working
according to global standards is especially important for
laboratories to ensure validity and global comparability of test and
calibration results. One of the goals of using global standards is
to reduce the number of tests required in national and international
trading.
The first edition of the “International
Standard General Requirements for the Competence of Testing and
Calibration Laboratories” was produced as a result of extensive
experience in implementing ISO/IEC Guide 25 and EN 45001; it
replaced these earlier standards in 1999. This new standard contains
all the requirements that testing and calibration laboratories have
to meet if they wish to demonstrate that they operate a management
system, are technically competent, and are able to generate
technically valid results.
Management requirements in the first edition
refer to ISO 9001:1994 and ISO 9002:1994. These standards have been
superseded by ISO 9001:2000, which made an update of ISO/IEC 17025
necessary. In the second edition of ISO/IEC 17025, released in 2005,
clauses were amended or added only when considered necessary in the
light of ISO 9001:2000
Testing and calibration laboratories that
comply with ISO/IEC 17025 will therefore also operate in accordance
with ISO 9001.
Accreditation bodies that recognize the
competence of testing and calibration laboratories use ISO/IEC 17025
as the basis for their accreditation.
ISO/IEC 17025 is divided into five clauses,
two annexes, and one bibliography section:
- Clause 1: Scope
The standard covers the technical activities of a laboratory as
well as the management and organizational aspects to perform the
technical activities in a competent way.
- Clause 2: Normative References
- Clause 3: Terms and Definitions
- Clause 4: Management Requirements
Most of the requirements are similar to those specified in the
ISO Standard 9001:2000.
- Clause 5: Technical Requirements
Most of the requirements come from the ISO Guide 25.
- Annex A: Cross References to ISO
9001:2000
- Annex B: Guidelines for Establishing
Applications for Specific Fields
- Bibliography
- The most important clauses are clause 4
and 5, describing management and technical requirements. In
addition to official requirements, these clauses also include
notes with further explanations and recommendations
Scope and Contents of this Tutorial
Implementing a quality system such as ISO/IEC
17025 has an impact on laboratory organization and operation. This
tutorial will discuss some of the specific implementation
requirements, along with their implications for testing and
calibration laboratories.
This tutorial is especially useful for
chemical analytical laboratories seeking accreditation according to
an internationally recognized standard. Examples include food
testing, environmental testing, chemical testing, clinical testing,
pharmaceutical testing, and other testing laboratories. This
tutorial will guide laboratory and QA managers and staff through the
entire process of ISO/IEC 17025 accreditation. It also helps
laboratories working under different quality systems to efficiently
set up procedures for compliance with all requirements.
Covered in this tutorial are:
- Management requirements
- Technical requirements
- Recommendations for Implementation
- Steps toward ISO/IEC 17025 accreditation
- Documentation
- Internal and external audits
- Implementing multiple quality systems
The tutorial and its reference material
should give a good understanding of the importance of ISO/IEC 17025,
the requirements, and the key points for implementation. The
tutorial is not a substitute for the standard itself and does not
list all requirements. Rather, it focuses on the most important
requirements and the ones that need specific attention, according to
the opinion and interpretation of the author. The tutorial
also does not include tools such as sample quality manuals,
operating procedures, and all the templates that would help to
quickly implement ISO/IEC 17025. These items can be obtained as
special packages that are available from service providers, for
example, the ISO/IEC 17025 Accreditation Package from LabCompliance
(2).
Quality Manual, Checklists and Procedures for
Easy Implementation
The Labcompliance ISO 17025 Accreditation
Package comes with an example Quality Manual, 2 checklists, 12 forms
and 39 SOPs for easy implementation. To learn more about the
package,
click here

Management requirements pertain to the
operation and effectiveness of the quality management system within
the laboratory. The requirements are similar to ISO 9001. This
clause is divided into fifteen chapters, described below.
Organization
This chapter ensures that the roles and
responsibilities of the laboratory, the management, and key
personnel are defined.
Key points:
- An organizational structure, as well as
responsibilities and tasks of both management and staff should
be defined.
- The organizational structure should be
such that departments having conflicting interests do not
adversely influence the laboratory’s work quality. Examples
include commercial marketing or financing departments.
- A quality assurance manager should be
appointed.
- All personnel should be free from any
commercial or financial pressure that could adversely impact the
quality of calibration and test results.
Management System
This chapter describes how to ensure that a
management system is implemented, maintained, and continually
improved.
Key points:
- There should be policies, standard
procedures and work instructions to ensure the quality of test
results.
- There should be a quality manual with
policy statements that are issued and communicated by top-level
management.
- The effectiveness of the management
system should be continually improved
Document Control
Individual paragraphs in this chapter describe
how to ensure that all documents related to the management system
are uniquely identified and created, approved, issued, and changed
following documented procedures.
Key points:
- All official documents should be
authorized and controlled.
- Documents should be regularly reviewed
and updated if necessary. The review
frequency depends on the document itself. Typical review cycles
are between one and three years.
- Changes to documents should follow the
same review process as for the
development of initial documents.
Review of Requests, Tenders, and Contracts
This chapter describes how to ensure that
requirements of requests, tenders and
contracts are well defined, reviewed, understood, and documented.
Key points:
- The laboratory supervisor’s review should
ensure that the laboratory has the technical capability and
resources to meet the requirements.
- Changes in a contract should follow the
same process as the initial contract.
Subcontracting of Tests and Calibrations.
This chapter describes how to ensure that
tests and calibrations subcontracted to third parties are performed
according to the same quality standards as if they were done in the
subcontracting laboratory.
Key points:
- The competence of the subcontracted party
should be ensured, through a documented quality system, such as
ISO/IEC 17025.
- The subcontracting laboratory is
responsible to the customer for the subcontractor’s work, except
in the case where the customer or the regulatory body specifies
which subcontractor should be used.
Purchasing Services and Supplies
This chapter describes how to
ensure that services and supplies delivered by third parties do not
adversely impact the quality and effectiveness of laboratory
operations.
Key points:
-
Suppliers should be
selected and formally evaluated to ensure that services and
supplies are of adequate quality.
-
Records of the selection
and evaluation process should be maintained.
-
The quality of incoming
material should be verified against predefined specifications.
Service to the Customer
This chapter describes how to
ensure that the laboratory continually meets customer requirements.
Key points:
-
The laboratory should
communicate with customers to clarify requests and get customer
input.
-
The laboratory should have
a formal program to collect feedback from customers on an
ongoing basis.
-
The laboratory should allow
customers to audit the laboratory.
Complaints
This chapter describes how to
ensure that any customer complaints are documented, evaluated, and
adequately followed up.
Key points:
-
There should be a policy
and procedure for the resolution of complaints received from
customers.
-
Records of complaints and
all steps taken when resolving the complaint should be
maintained. This includes documentation of investigations and
corrective actions.
Control of Nonconforming Testing and/or
Calibration Work
Tests, calibrations, and other
laboratory operations should conform to previously
defined specifications such as laboratory specifications or
client-defined specifications.
This chapter describes how to
ensure that nonconforming test and calibration results are
adequately followed up, and that corrections are initiated.
Key points:
-
There should be a policy
and process that come into effect when results do not conform to
procedures.
-
Corrective actions should
be taken immediately to avoid recurrence.
-
The significance of
nonconforming work should be evaluated, for example, the
possible impact on other testing or calibration work.
-
If necessary, customers
should be notified.
Continuous
Improvement
This chapter describes how to
ensure that the effectiveness of the management
system is continually improved.
Key points:
-
Suggestions for
improvements should be taken from audit reports, analysis of
data, customer complaints and suggestions, corrective and
preventive actions, and management reviews.
-
Suggestions should be
collected over time and reviewed by management for suitable
actions.
Corrective Action
This chapter describes how to
ensure that the root cause of nonconforming work
or deviations from laboratory and management procedures are
identified and that
adequate corrective actions are selected, implemented, documented,
and monitored.
Key points:
-
Corrective actions can be
triggered through nonconforming tests or other work, customer
complaints, internal or external audits, management reviews, and
observations by staff.
-
Corrective actions should
be selected and implemented to eliminate the specific problem
and prevent recurrence of the same problem.
-
As the first step in the
process, the root cause of the nonconformity should be
identified.
-
The effectiveness of the
corrective action should be monitored and evaluated.
Preventive Action
Preventive actions should be
initiated when potential sources of nonconformities
have been identified. Nonconformities may be technical or related to
the management system. The objective is to reduce the likelihood of
the occurrence of such potential nonconformities.
Key points:
-
There should be a procedure
to identify potential sources of nonconformities and define
preventive actions to prevent occurrence of these
nonconformities.
-
The effectiveness of the
preventive action should be monitored and evaluated.
Control of Records
This chapter describes how to
ensure that all records in a laboratory are uniquely
identified, readily available when needed, and protected against
unauthorized access for viewing or changing.
Key points:
-
There should be procedures
for identification, collection, indexing, storage, retrieval,
and disposal of records
-
Records should be stored
such that their security, confidentiality, quality and integrity
are ensured throughout the required retention time
-
For technical records such
as test reports of analytical measurements, original
observations should be retained, along with processing
parameters that will allow tracking final results back to the
original observations.
-
Record format can be hard
copies or electronic media. There should be procedures to
protect and back-up electronic records and to prevent
unauthorized access.
-
Records can be corrected if
there are mistakes. The original record should be crossed out,
but still visible.
-
When electronic record
systems are used, the same principle applies. The laboratory
should ensure that original records are not overwritten by the
system and that corrections are recorded together with the
original records. Using a system that prevents overwriting
original records and stores changes in an electronic audit trail
that can be viewed and printed is highly recommended.
Internal Audits
Internal audits should verify
that the laboratory complies with ISO/IEC 17025 and with internal
technical and quality procedures. Internal audits are also an
excellent preparation for external assessments and can help to
continually improve the quality system.
Key points:
-
The laboratory should have
a procedure and a schedule for internal audits. Internal audits
can either cover the whole laboratory and all elements of the
quality system at one specific period of time or can be divided
into several subsections.
-
The schedule should be such
that each element of the quality system and each section of the
laboratory are audited yearly.
-
The audit program should be
managed by the quality manager.
-
Audit findings related to
the quality of test and calibration results should be reported
to customers.
-
Audit follow-up activities should include
corrective and preventive action plans(CAPA). The effectiveness
of the plans should be monitored.
Management Reviews
Requirements in this chapter
describe how to ensure the continued suitability and
effectiveness of the quality system, policies, and testing and
calibration procedures.
Key points:
-
There should be a schedule
and procedure for periodic management reviews. Recommended
review frequency is once a year.
-
The management review
should include a discussion about the outcome of recent internal
audits and external assessments, corrective and preventive
actions, results of proficiency testing, customer complaints and
feedback, and any recommendations for improvements.
-
Management should decide on
follow-up activities. These activities should be monitored for
effectiveness.
Procedures to Implement Management
Requirements
The Labcompliance ISO 17025 Accreditation
Package comes with a full set of procedures to easily implement
management requirements. To learn more about the package,
click
here
Technical requirements address
the competence of staff, sampling and testing methodology,
equipment, and the quality and reporting of test and calibration
results.
This clause is divided into ten
chapters.
General
The technical requirements
clause starts with a general chapter. This chapter’s purpose is to
make readers aware that the correctness and reliability of test and
calibration results are determined by a variety of factors
Key points are:
- The different factors impacting the
quality of results should be documented. They include, for
example, sampling, equipment, test methods, and environmental
conditions.
- The extent to which impacting factors can
contribute to the measurement uncertainty should be taken into
account when developing test and calibration methods..
Personnel
Personnel probably have the highest impact on
the quality of test and calibration
results. This chapter describes how to ensure that all laboratory
personnel who can impact test and calibration results are adequately
qualified.
Key points are:
- Only competent personnel should perform
testing and calibrations. This includes part- time as well as
full-time employees, as well as all management levels
- Competence can come from education,
experience, or training.
- Management should define and maintain
tasks, job descriptions, and required skills for each job.
- Based on required skills and available
qualifications, a training program should be developed and
implemented for each employee.
- The effectiveness of the training should
be evaluated. If the training is related to a specific test
method, the trainee can demonstrate adequate qualification
through successfully running a quality control or proficiency
test sample. A statement from the trainee such as ‘I have read
through the test procedure’ is not enough.
- Management should authorize personnel to
perform specific tasks, for example, to operate specific types
of instruments, to issue test reports, to interpret specific
test results, and to train or supervise other personnel.
- The date of this authorization should be
recorded. The associated tasks should not be performed before
the authorization date.
Accommodation and Environmental Conditions
This chapter has been included to ensure that
the calibration and test area environment will not adversely affect
the measurement accuracy. It includes five sections with information
that is mostly common sense. One clause recommends having effective
separation between neighboring areas when the activities therein are
incompatible. An example would be to separate laboratories that
analyze extremely low traces of a solvent from those which consume
large quantities of the same solvent for liquid-liquid extraction.
Key points are:
- Environmental conditions should not
adversely affect the required quality of tests. This means, for
example, that equipment should operate within the manufacturer’s
specifications for humidity and temperature.
- The laboratory should monitor, control,
and record environmental conditions. Special attention should be
paid to biologic sterility, dust, electromagnetic disturbances,
radiation, humidity, electrical supply, temperature, sound, and
vibration. Tests should be stopped when the environmental
conditions are outside specified ranges.
- Areas with incompatible activities should
be separated.
- Access to test and calibration areas
should be limited to authorized people. This can be achieved
through pass cards..
Test and Calibration Methods and Method
Validation
Accurate test and calibration results can only
be obtained with appropriate methods that are validated for the
intended use. This chapter deals with the selection and validation
of laboratory-developed and standard methods and measurement
uncertainty and control of data.
Key points for accurate test and calibration
results:
- Methods and procedures should be used
within their scope. This means the scope should be clearly
defined.
- The laboratory should have up-to-date
instructions on the use of methods and equipment.
- If standard methods are available for a
specific sample test, the most recent edition should be used.
- Deviations from standard methods or from
otherwise agreed-upon methods should be reported to the customer
and their agreement obtained.
- When using standard methods, the
laboratory should verify its competence to successfully run the
standard method. This can be achieved through repeating one or
two critical validation experiments, and/or through running
method specific quality control and/or proficiency test samples.
- Standard methods should also be validated
if they are partly or fully out of the scope of the test
requirement.
- Methods as published in literature or
developed by the laboratory can be used, but should be fully
validated. Clients should be informed and agree to the selected
method.
- Introduction of laboratory-developed
methods should proceed according to a plan.
- The following parameters should be
considered for validating in-house developed methods: limit of
detection, limit of quantitation, accuracy, selectivity,
linearity,
repeatability and/or reproducibility, robustness, and linearity.
- Exact validation experiments should be
relevant to samples and required information.
- Sometimes, standard and in-house
validated methods need to be adjusted or changed to ensure
continuing performance. For example, the pH of a HPLC mobile
phase may have to be changed to achieve the required separation
of chromatographic peaks. In this case, the influence of such
changes should be documented, and if appropriate, a new
validation should be carried out.
- Validation includes specification of the
requirements and scope, determination of the characteristics of
the methods, appropriate testing to prove that the requirements
can be fulfilled by using the method, and a statement on
validity.
Key points for measurement uncertainty:
- The laboratory should have a procedure to
estimate the uncertainty of measurement for calibrations and
testing.
- For uncertainty estimation the laboratory
should identify all the components of uncertainty.
- Sources contributing to the uncertainty
can include the reference materials used, the methods and
equipment used for sampling and testing, environmental
conditions and personnel.
Key points for control of data:
- Calculations used for data evaluation
should be checked. This is best done during software and
computer system validation. As an example, spreadsheet formulas
defined by a specific user should be verified with an
independent device such as a handheld calculator. Data transfer
accuracy should be checked. Accuracy of data transfer between
computers can be automatically checked with MD5 hash sums.
- Computer software used for instrument
control, data acquisition, processing,
reporting, data transfer, archiving, and retrieval developed by
or for a specific user should be validated. The suitability of
the complete computer system for the intended use should also be
validated.
- Any modification or configuration of a
commercial computer system should be validated. Examples include
defining report layouts, setting up IP addresses of network
devices, and selecting parameters from a drop-down menu.
- Electronic data should be protected to
ensure integrity and confidentiality of
electronic records. For example, computers and electronic media
should be
maintained under environmental and operating conditions to
ensure integrity of data.
Equipment
Equipment that is performing
well and properly maintained is a prerequisite for the ongoing
accuracy of test and calibration results. This chapter deals with
the capacity and quality of equipment. The whole idea is to make
sure that the instrument is suitable for performing selected
tests/calibrations and is well characterized, calibrated, and
maintained.
Key points are:
- Equipment should conform to
specifications relevant to the tests. This means that equipment
specifications should first be defined so that when conforming
to defined specifications the equipment is suitable to perform
the tests.
- Equipment and its software should be
identified and documented.
- Equipment should be calibrated and/or
checked to establish that it meets the laboratory’s
specification requirements.
- Records of equipment and its software
should be maintained and updated if
necessary. This includes version numbers of firmware and
software. It also includes calibration and test protocols.
- Calibration status should be indicated on
the instrument along with the last and the next calibration
dates..
Measurement Traceability
Traceability of equipment to the same standard
is a prerequisite for comparability of test and calibration results.
Ideally all measurements should be traceable to International System
of Units (SI). While this is typically possible for physical
measurements such as length (m) and weight (kg), this is more
difficult in chemical measurements.
Key points for traceability of calibrations:
- Calibration of equipment should be
traceable to the SI units.
- Traceability of laboratory standards to
SI may be achieved through an unbroken link of calibration
comparisons between the laboratory standard, secondary standard,
and primary or national standard.
- If traceability to SI units is not
possible, the laboratory should use other appropriate
traceability standards. These include the use of certified
reference material and the use of consensus standards or
methods.
Sampling
This chapter has been added to describe how to
ensure that statistically relevant
representative samples are taken and that all information on the
sample and the
sampling procedure is recorded and documented.
Key points for sampling:
- Sampling should follow a documented
sampling plan and sampling procedure.
- The sampling plan should be based on
statistical methods.
- The sampling procedure should describe
the selection and withdrawal of representative samples.
- The sampling location and procedure, the
person who took the sample, and any other relevant information
about the sampling process should be recorded.
Handling Test and Calibration Items
This chapter describes how to
ensure that sample integrity is maintained during transport,
storage, and retention and that samples are disposed of safely.
Key points for handling test and calibration
items are:
- Test and calibration items should be
uniquely identified.
- Sample transportation, receipt, handling,
protection, storage, retention, and/or disposal should follow
documented procedures.
- The procedures should prevent sample
deterioration and cross-contamination during storage and
transport.
Assuring the Quality of Test and Calibration
Results
This chapter describes how to ensure the
quality of results on an ongoing basis through, for example, regular
analysis of quality control samples or participation of
proficiency-testing programs.
Key points are:
- The validity of test results should be
monitored on an ongoing basis.
- The type and frequency of tests should be
planned, justified, documented, and reviewed.
- Quality control checks can include the
regular use of certified reference materials, replicating tests
or calibrations using the same or different methods, and
retesting or recalibration of retained items.
Reporting of Results
This chapter describes how test/calibration
results should be reported. This is important for an easy comparison
of tests performed in different laboratories. The chapter has some
general requirements on test reports such as clarity and accuracy,
but it also has very detailed requirements on the contents..
Test reports and calibration certificates
should include:
- The name and address of the laboratory.
- Unique identification of the test report
or calibration certificate (such as the serial number).
- The name and address of the client.
- Identification of the method.
- A description and identification of the
item(s) tested or calibrated.
- Reference to the sampling plan and
procedures used by the laboratory.
- The test or calibration results with the
units of measurement.
- The name(s), function(s) and signature(s)
or equivalent identification of person(s) authorizing the test
report or calibration certificate.
- A statement on estimated uncertainty of
measurement (for test reports ‘where applicable’).
- When opinions and interpretations are
included, documentation of the basis for the opinions and
interpretations.
- Opinions and interpretations clearly
marked as such on the test report or calibration certificate
Procedures to Implement Technical
Requirements
The Labcompliance ISO 17025 Accreditation
Package comes with a full set of procedures to easily implement
technical requirements. To learn more about the package,
click here
Now that you have an overview of required management and technical
controls, we will give recommendations on how to efficiently comply
with some key requirements. The scope of this primer does not go
into all the details and we cannot cover all requirements.
We will focus on topics that most likely are
new to laboratories without a quality system. These include:
specific organizational structure, formal equipment calibration and
testing, measurement traceability and uncertainty. We try to balance
the lack of detailed information by providing references to official
guidelines, textbooks, and other literature. For example,
EURACHEM/CITAC, EUROLAB and ILAC have developed guidance documents
for measurement uncertainty and traceability in measurement (3-9).
ISO has published a “Guide to the Expression of Uncertainty in
Measurement” (10). A EURACHEM-UK working group has developed
guidance for qualification of analytical instruments that was
published in “Accreditation and Quality Assurance” (11). EUROLAB has
published a technical report called “Management of Computers and
Software in Laboratories with Reference to ISO/IEC 17025:2005” (12).
Huber has authored a textbook “Validation and qualification in
Analytical Laboratories” and Thompson et al. gave recommendations
for an “International Harmonized Protocol for Proficiency Testing of
Chemical Analytical Laboratories” (14).
Help is also readily available from
accreditation bodies. For example, A2LA has a “Policy on Measurement
Traceability” (15) and LabCompliance provides a complete “ISO/IEC
Accreditation Package” with a sample master plan, SOPs, forms, and
checklists (2).
Organizational Structure
ISO/IEC requires that organizational arrangements should be such
that departments with conflicting interests do not adversely
influence compliance with the standard. For example, finance and QA
should operate independently from laboratory activities. Figure 2
gives an example of how this could be accomplished. Finance and QA
do not report to laboratory management but rather to the director of
the company.

Figure 2: Example for Organizational Structure
(from Reference 2)
Equipment
Each laboratory should have a plan for how to
ensure adequate equipment function and performance before and during
sample measurement. The main activities are calibration and checking
to verify specified performance and maintenance. The extent of
testing depends on the complexity and use of the equipment. Each
laboratory should have processes for how calibration and testing is
performed for different types of equipment.
Preferably the whole range of equipment should
be divided into a few categories such as A, B, and C, and
calibration and/or verification tests should be associated with each
category. Instruments in the simplest category (A) such as stirrers
and mixers usually don’t need to be tested but only visually
inspected. Category B instruments such as balances and pH meters
should be calibrated according to manufacturer SOPs, and more
complex instruments such as chromatography systems should be fully
tested according to their intended use.
The following paragraphs contain
recommendations for how to specify, test, and
maintain analytical equipment for ISO/IEC 17025 compliance.
Documenting Specifications
ISO/IEC 17025 requires that equipment and
software should comply with specifications that are relevant to the
tests. Therefore, the first step in the process is to define and
document the equipment specifications.
- For simple equipment, such as balances
and pH meters, the use of manufacturer specifications is
recommended.
- For more complex equipment hardware, such
as gas chromatographs or mass
spectrometers, the manufacturer’s specifications can also be
used. This is only recommended as long as all vendor-specified
functions are required by the intended applications over the
fully specified range. As an alternative, the user can define
specifications according to the intended use of the instrument.
- Commercial software and computer systems
typically provide more functionality than required by a specific
user. Therefore, for computer systems, the user should define
specifications according to the system use. A functional
specifications list will help
define user specifications.
Selecting a Vendor
A documented procedure and
well-defined criteria should be used for selecting
equipment suppliers. Useful criteria include:
-
Vendor’s equipment meets
the user’s requirement specifications.
-
Vendor has leading position
in the marketplace.
-
Equipment and software
design, development, and manufacturing takes place in a quality
system environment such as ISO 9001.
-
Vendor provides
installation, familiarization, and training services.
-
Metrology-based calibration
and functional testing services are performed through qualified
engineers.
-
Vendor provides phone and
onsite support in local language.
Installation and Documentation
Installation can be performed
by the vendor or by the user. Steps include:
-
Verify that the location
meets the environmental specifications as defined by the vendor.
-
Install equipment hardware
according to vendor specifications.
-
Install software and
start-up according to vendor specifications.
-
Create documentation of
hardware and software, such as vendor, product number, model
number, serial number, and location.
Initial Testing for Calibration and/ or
Performance Verification
Equipment used for measurement
should be tested before initial use to ensure acceptable
performance. This is performed through calibration such as the mass
of a balance, or through verification of specified performance
characteristics such as the sampling precision of a gas
chromatograph.
Steps for testing include:
-
Develop test procedures and
test protocols.
-
Define acceptance criteria
based on documented specifications.
-
Select and order traceable
test tools, such as reference weights for calibration of a
balance.
-
Make sure that test
engineers have the appropriate qualifications to perform the
tests.
-
Perform the tests and
document test results.
-
Verify that acceptance
criteria are met.
-
Label the equipment with
the status, as well as the dates of last and next calibrations.
-
Maintain records of
calibration and checks.
These tests can be performed by
the vendor or by the user. Advantages of testing by the vendor can
be demonstrated using the Agilent Functional Verification Service
(FVS) for gas and high-performance liquid chromatography. This
service was specifically developed to meet ISO/IEC 17025
requirements:
-
Agilent has decades of
knowledge and experience with factory and field-testing of
equipment. As a result, the test selection and sequence is
optimized for the highest speed and lowest instrument downtime,
without comprising accuracy or calibration specifications.
-
Agilent engineers bring
along calibrated test tools that meet traceability requirements.
-
Agilent engineers also
bring certificates to document qualification.
-
Agilent tests are combined
with recommended preventive maintenance for all the critical
functional components of the equipment.
-
Agilent provides global
lab-to-lab consistency via standardized and well-recognized
maintenance and verification testing protocols.
-
Agilent’s calibration and
test certificates are globally recognized by internal auditors
and official assessors.
Testing During Ongoing Use
Equipment characteristics and performance can
change over time. Therefore, equipment quality programs should
ensure that equipment is routinely tested on an ongoing basis.
The type and frequency of such tests depends
on the equipment. For example, on a daily basis, balances can be
checked with laboratory reference weights, and chromatographs can be
checked using well-defined quality control samples.
In addition to the more frequent tests that
only challenge a subset of all specifications, annual repetition of
all initial tests is recommended to repeat all the initial tests as
described above, for example, for chromatographs, once a year. A
balance is also typically calibrated once a year by the vendor with
calibrated and traceable standard weights.
Maintenance and Repair
The laboratory should have a maintenance plan
to carry out preventive maintenance activities and a procedure
for unplanned repair to ensure ongoing performance and
reliability.
- Defective equipment should not be used
for tests and calibration. Smaller devices should be taken out
of the laboratory and bigger instruments should be clearly
labeled as being defective. Specified functioning and
performance should be verified after repair.
- Records of maintenance and repairs should
be maintained.
Software and Computer Systems
ISO/IEC 17025 requires computer systems and
software used for acquisition, processing, recording, reporting,
storage and retrieval of test and calibration data be validated when
the software is developed, configured, or customized by the user.
EUROLAB has developed a technical report with
recommendations for “Management of Computers and Software in
Laboratories with Reference to ISO/IEC 17025:2005” (12).
This report provides advice on validation steps for different
software and system risk categories, as well as recommendations for
how to ensure the security, availability, integrity, and
confidentiality of electronic records.
The report divides software into five
categories as listed in Figure 3 and described below::
- Operating systems
- Firmware as built into automated
equipment
- Standard software packages such as word
processors and non-configurable
computerized analytical systems
- Configured software packages such as
Excel formulas and configurable
computerized analytical systems
- Custom built software

Figure 3: Validation Activities for Software
Categories (Simplified).
Measurement Traceability
ISO/IEC 17025 requires reference material used
for calibration of measurement
equipment to be traceable to SI Units, where possible. Typically,
laboratories use their own internal reference material for
calibration. Traceability of such material to SI units can be
achieved through an unbroken chain of comparisons between laboratory
reference material and SI units. An example is shown in Figure 4.

Figure 4: Traceability Chain of Laboratory
Reference Material
Working standards are regularly compared with secondary standards by
an accredited laboratory These secondary standards are
calibrated by a national metrology institute or an accredited
reference laboratory. For this kind of comparison, the measurement
uncertainty should be known and documented in calibration
certificates so that the measurement uncertainty of the working
standard can be estimated and reported.
This concept works well for physical
measurement such as meter (m) for length, kilogram (kg) for mass,
and Kelvin (K) for temperature. For this reason, reference balance
masses, thermometers and thermocouples should be traceable to SI
Units through an unbroken chain of comparisons performed by
accredited laboratories and national metrology institutes..
For most reference material
used for chemical measurements, traceability to SI units is very
difficult and not practical. The traceability chain in Figure 4 ends
at the lower level, at suppliers of standard reference material
(NIST, for example), at suppliers of certified reference material,
or at a company’s accredited metrology laboratory. When traceability
to SI is not possible, ISO/IEC 17025 recommends the use of
well-characterized certified reference material that is provided by
a competent supplier. Alternative well-defined methods (also called
primary or definitive methods) agreed upon by all parties can also
be used to establish traceability. This topic has been discussed in
different working groups. Detailed recommendations have been
published by ILAC (7) and EURACHEM/CITAC (8).
X.R. Pan (17) suggested a
classification scheme of reference material used for chemical
measurements. The classification as described on the following page
is well accepted in chemical laboratories.
Primary Reference Material
- Also called primary standards.
- Developed by a national metrology
laboratory.
- Certified by primary/consensus method.
- Traced back to SI units and/or verified
by international comparison
Certified Reference Material
- Also called secondary standards.
- Derived from primary reference material
with statement of uncertainty.
- Usually prepared by a specialized
reference laboratory.
- Certified by reference methods or
comparison methods.
- Recognized by national or otherwise
specialized authoritative organization.
Working Reference Material
- Also called internal reference material.
- Derived from certified reference
material.
- Accuracy verified by well characterized
and validated methods.

Measurement Uncertainty
There is uncertainty associated
with every test and calibration. For testing, this is due to errors
arising at the various stages of sampling, sample preparation,
measurement, and data evaluation. In other words, whenever any
quantitative measurement is performed, the value obtained is only an
approximation of the true value. Users of the measurement data
should have an idea of how much the reported result may deviate from
the true value. ISO/IEC 17025 recommends reporting the results of
quantitative measurement both as a single value and with the
possible deviation from the true value. This is logical for any
report with quantitative results. It is, for example, of no use if a
report on a food sample states 0.1 percent of compound X, and the
user of the data is still unsure whether this could be 0.05 or 0.4
percent due to the level of uncertainty.
An uncertainty statement
provides the user with information on the approximate measurement
tolerances and the expected limits within which the true value of
the measurement, such as analyte concentration, is supposed to lie.
Without documenting the uncertainty, although the analyst can
estimate the level of uncertainty, the client or user of the data
cannot.
Information on uncertainty is
of particular importance if a specification limit is to be verified
and reported. For example, if a purchasing agreement specifies that
a product can only be released if compound X is below 0.5 percent,
the test report may not contain a statement about compliance if the
measurement results range including the measurement uncertainty is
above 0.5 percent.
When parameters are claimed to
be within a specified tolerance, the measurement values range,
including the estimated uncertainty of measurement, shall fall
within the specification limit.
ISO has published a “Guide to
the Expression of Uncertainty in Measurement” (10). It establishes
general rules for evaluating and expressing uncertainty in
measurement across a broad spectrum of measurements.
EURACHEM has produced an
excellent document containing great detail about how the concepts of
the ISO guide can be applied in chemical measurement (4). The whole
process of measurement uncertainty is schematically shown in Figure
5. The basic ideas are explained in this primer, but for more
detailed information, readers are encouraged to study the EURACHEM
document (4).
The concept of evaluating
uncertainty is fairly straightforward. It requires a detailed
knowledge of the nature of the items being measured and of the
measurement method, rather than an in-depth understanding of
statistics.

Figure 5: Estimating Measurement Uncertainty
- Develop the specifications by writing a
clear statement of exactly what is to be measured and the
relationship between this and the parameters on which it
depends. For example, if the measurement temperature has an
influence on the result, the measurement temperature should also
be defined.
- Develop a workflow diagram for the
entire, sample collection, sample preparation, calibration,
measurement, data evaluation, and data transcription process
(see Figure 1 for analytical sample testing).
- Identify and list sources of uncertainty
for each part of the process or for each parameter. Possible
sources for errors may be derived from non-representative
sampling, operator bias, a wrongly calibrated instrument, lack
of ideal measurement conditions, chemicals with impurities, and
errors in data evaluation.
- Estimate and document the size of each
uncertainty, for example, as standard
deviations or as RSDs. The data should be gathered from a series
of measurements. Where experimental evaluation is impossible or
impractical, the individual contributions should be estimated
from whatever sources are available. Sources for this kind of
estimation can be found in the supplier’s information or in the
results of inter-laboratory studies or proficiency testing. The
procedures and thoughts behind the way the contributions have
been measured or estimated should be documented.
- Combine separate contributions in order
to give an overall value. For example, where individual sources
of uncertainty are independent, the overall uncertainty can be
calculated as a multiple of the sum of squared contributing
uncertainty components, all expressed as standard deviations.
Computer software or spreadsheet programs can help automate this
calculation.
The whole procedure should be documented in
such a way that sufficient information is available for the result
to be reevaluated if new information or data become available.
A complete set of documentation should include:
- A description of the methods used to
calculate the measurement result and its
uncertainty from the experimental measurements
- The values and sources of all corrections
- A list of all components of uncertainty
with full documentation about how each
of these was evaluated.
Reference 4 includes many practical examples
with data from different analyses, as well as formulas for
evaluating, calculating and reporting standard and expanded
uncertainty. Sample analysis reports should include an uncertainty
number, which is typically expressed as:
Result = x ± u (units)
or
Result = x (units)
Uncertainty = u (units)
Implementing Difficult Requirements
The Labcompliance ISO 17025 Accreditation
Package comes with procedures and forms to easily implement and
document difficult requirements. They include measurement
uncertainty, traceability, validation of spreadsheets and training.
To learn more about the package,
click here
ISO/IEC 17025 accreditation should be
carefully thought out and properly prepared. It can be quite
expensive but can also have big benefits. The balance between costs
and benefits should be calculated and documented. Implementing
ISO/IEC 17025 will impact the entire laboratory and also supporting
services such as human resources, documentation and finance
departments. Therefore, while the decision to initiate and fund the
project will be made by management, all affected departments should
be involved in the process. The entire process is divided into two
phases: investigation phase and implementation phase. Figure 6
illustrates the steps for both phases.

Figure 5. Steps Towards ISO/IEC 17025
Accreditation
Investigation
In the investigation phase, information is
collected to decide if becoming accredited makes good business sense
and typically follows these steps:
- Management initiates, funds, and
otherwise supports the investigation.
- Management nominates a project owner.
Ideally the person should have
experience in laboratory operations, good business sense, an
understanding of
quality systems and excellent communication skills.
- The project manager with support from
management recruits a project team.
Members should come from laboratory management, QA, finance,
human
resources, training, and documentation groups.
- The project team defines the scope of the
intended accreditation. This could
include all calibrations and/or tests performed in a lab, or
just part of them.
- The project team studies the
accreditation requirements in detail. The main source is the
standard ISO/IEC 17025, with additional support from official
guidelines, other literature, and external expert advice.
- The project team develops a requirements
list. The list should include all documents as required by the
standard, for example, policies, a quality plan, and procedures
for most of the requirements.
- The project team prepares a gap analysis
by comparing ISO requirements as listed above with what is
already available and implemented. A gap exists where existing
policies, processes or procedures do not fully meet the stated
requirements. This analysis should include all processes and
procedures for management controls and technical controls, such
as for sampling, method validation, equipment calibration,
qualification and maintenance, employee qualifications, and
others.
- Using the outcome of the gap analysis,
the project team develops a task list. The list is completed
with additional tasks such as selecting and dealing with an
accreditation body.
- The project team, together with the help
of an external consultant, makes an estimation of the overall
ISO/IEC implementation costs, which should include costs for
initial set-up and also for maintaining the quality system. The
costs are compared with the direct and tangible estimated
additional returns that come from getting accreditation status.
Tangible returns are, for example, savings through more
efficient operation.
- The team makes a rough estimation of the
return on investment for both the short and long term views, and
makes a recommendation to management.
- Management decides to accept or reject
the proposal and whether to proceed with accreditation.
Implementation
Once the decision for ISO/IEC 17025
accreditation is made, the laboratory develops
and implements documentation in preparation for the accreditation
assessment.
Typically implementation follows these steps:
- The project owner forms implementation
teams for different areas. It is critical that all affected
departments at all management levels are represented in the
teams.
- The project owner with the help of QA
searches for an accreditation body and selects the one that best
fits the laboratory’s needs. There are several ways to find
accreditation bodies. Probably the best way is to ask other
accredited laboratories about their experiences.
- The teams develop documentation such as
procedures under the supervision
of the project owner.
- The project owner arranges for staff
training.
- Quality assurance performs an internal
audit and initiates corrective actions,
if necessary.
- The selected accreditation company
performs a pre-assessment.
- The project owner initiates corrective
actions.
- The accreditation company performs an
accreditation audit.
ISO 17025 Implementation: Step-by-Step
The Labcompliance ISO 17025 Accreditation
Package comes with a best practice guide for effective step-by-step
implementation of ISO 17025 - from beginning to the end!. To learn
more about the package,
click here
ISO/IEC 17025 requires
different types of documentation, as illustrated in the
documentation pyramid in Figure 7.

Figure 7: Documentation Pyramid
A policy documents the laboratory’s intent to
implement ISO/IEC 17025. The Quality Manual is the top tier of the
document hierarchy. It describes the approaches to achieve quality
data. It also includes policy statements describing the laboratory’s
intention to conform to ISO/IEC 17025 requirements. For example, a
policy statement could be: All personnel involved in calibration and
testing should be competent for the assigned task.
A process or generic procedure describes how
various quality requirements can be
achieved. For example, it describes how the requirement ‘Personnel
should be
competent for the assigned task’ can be implemented.
Standard operating procedures (SOPs) or
Working Procedures are step-by-step instructions for how to exactly
perform a specific task, such as calibrating a specific instrument.
Records are generated on a day-by-day basis,
such as analytical results from product tests or calibration records
of a balance.
All documents should be properly controlled.
For example, each change should be
authorized and logged, and the updated document should get a new
revision number or code.
Policies and Quality Manual
Policies including the quality policy
statement should be documented in the Quality Manual. Senior
management should write the quality policy statement, which should
outline the laboratory’s commitment to quality. The quality manual
describes the quality system and documents the laboratory’s goal and
overall concept for how to conform to ISO/IEC 17025. It should also
describe how the remainder of the quality system documentation is
organized. It should be developed by working groups representing
different departments.
Processes
Processes or standard procedures describe how
various ISO/IEC 17025 requirements can be achieved. For example, it
describes how the requirement ‘All personnel involved in calibration
and testing should be competent for the assigned task’ can be
implemented. Another example is the laboratory’s approach to
calibrating and checking different types of equipment. For a better
understanding, process flowcharts should be included in a process
description.
Standard Operating Procedures (SOPs) and Work
Instructions
Routine activities follow documented
procedures. These are typically defined as standard operating
procedures (SOPs) or work instructions. While quality manuals and
processes describe tasks and approaches, procedures and work
instructions give step-by-step instructions on performing tasks.
Examples of SOPs are procedures for checking and calibration of
equipment. All laboratory SOPs should use the same format, to make
writing and reading easier. A good practice is to have an SOP for
how to author, review, approve, distribute, and update SOPs.
Preferably senior members of anticipated user groups should write
SOPs. This helps ensure that SOPs have the right level of
information and are used and followed.
Records
Records to demonstrate conformity with ISO/IEC
17025 and as required by customers should be retained for a specific
amount of time. Examples are original laboratory observations, test
results, supporting documents such as chromatograms, and training
certificates and equipment calibration protocols.
Checklists, forms, templates, and examples
help implement quality work effectively and consistently. Examples
of these include checklists and worksheets for vendor assessment,
handling nonconforming test results, and for internal audits. These
items help document specific tasks consistently and effectively.
Document Control
Development and maintenance of documentation
should be controlled through document control and management
procedures that are part of the management system. Documents include
both internal, such as SOPs, quality manuals, and training plans and
external documents, such as regulations, standards, test methods,
and instrument operating manuals.
The procedure for document control should
ensure that:
- Official documents are created or
acquired, reviewed, and approved prior to use.
- Documents are uniquely identified with
document and revision number, date of revision, and issuing
authority.
- A quality list with all controlled
documents is maintained by QA. The list includes document and
revision number, title, date of issue, date of last review, and
locations.
- Internal documents include page numbers
and total number of pages on each page. Users of the documents
are adequately trained before the documents are released.
- Current authorized versions of documents
are readily available at the user’s workspace.
Quality managers manage internal audits. They
verify conformance to the ISO/IEC
17025 requirements and also to company policies, processes and
procedures. Internal audits are also quite useful in preparation for
external audits. External auditors can come from clients or from
accreditation bodies. They verify that the laboratory is operating
in compliance with ISO/IEC 17025.
There should be procedures for staff
responsibilities before, during, and after internal and external
audits. Overall owners should be defined, and all employees who may
be affected by the audit should be trained. This chapter summarizes
recommendations for audits. To make best use of internal audits,
they should be designed, executed, and followed up very much in the
same way as expected external audits. The recommendations for audit
preparation, performance, documentation, and follow-up are written
for the audited departments, not for the auditors.
Internal Audit Schedule
Internal auditing should follow a
predetermined schedule covering all activities over a reasonable
period of time. It is inconvenient to audit all activities in a
single audit, so it should be spread over several quarterly or
monthly audits. The schedule for such audits is conveniently drawn
as a matrix covering, for example, a year in which dates are set for
each part of the quality system. Audit schedules can be organized as
horizontal or vertical. A vertical audit checks compliance of, for
example, a single test through all steps from sampling to archiving
of records. A horizontal audit examines every aspect of a single
requirement, for example, equipment. Figure 8 shows an example of a
horizontal audit..

Figure 8: Example for Horizontal Audit
Schedule
Audit Phases
Internal audit activities are spread over
different phases which included preparation, conduct, close and
follow-up. Typical steps for each phase are listed below:
Preparation
- Assign an overall owner and host for the
audit
- Assign a technical contact to get access
to and review the completeness of records and other
documentation for items to be audited. The assigned technical
contact should be present all the time.
- Set up a work area for the inspectors.
- Review the schedule.
- Prepare and train staff.
- An audit may be a tough experience for
all people involved, so everyone needs to be informed about what
will happen and what questions may be asked.
Conduct
- Maintain a continuous log of the audit.
- Provide copies (do not give originals
away).
- Keep duplicates of all information
supplied to auditors.
- Take immediate corrective action, when
appropriate.
- Hold a daily debriefing meeting to assess
the progress.
- Keep all documents in the work area.
- Accompany the auditor at all times.
- Be courteous and cooperative.
- Answer only questions that are asked.
- If you are unable to answer, tell the
auditor openly.
- Protect proprietary information.
Close
- Clarify any open questions or causes for
dissatisfaction in the exit meeting
Follow-up
- Develop a corrective action plan (owners,
tasks, deliverables, and schedule).
- Develop a preventive action plan (owners,
tasks, deliverables, and schedule).
- Monitor the plan
SOP and Checklists for Internal Audits
The Labcompliance ISO 17025 Accreditation
Package comes with a procedure and checklists for internal
audits. They are optimized to serve as learning esperienience for
external assessments. To learn more about the package,
click here
Laboratories are frequently faced with situations where they have to
comply with both regulations and quality standards at the same time.
Examples are:
- A clinical laboratory performs contract
analyses from pre-clinical and clinical studies for
pharmaceutical companies. The laboratory also performs special
tests for hospitals. The laboratory has to operate in compliance
with U.S. FDA and EU GLP and GCP regulations for clinical and
pre-clinical study tests. Some customers also require laboratory
accreditation according to ISO/IEC 17025, others according to
ISO 15189, a standard for medical laboratories.
- A chemical company is certified for ISO
9001. The scope of the certification also covers the analytical
service laboratory. In addition, the laboratory performs
contract analyses for other companies and has received
laboratory accreditation in compliance with ISO/IEC 17025. The
laboratory has to work in compliance with ISO 9001 and with
ISO/IEC 17025.
- An independent test laboratory performs
GLP studies as a subcontractor for a pharmaceutical company.
Occasionally, the laboratory also performs analyses for
pharmaceutical manufacturing control departments. The laboratory
has also received laboratory accreditation for specific food
analyses according to ISO/IEC 17025. The laboratory has to
comply with ISO/IEC 17025 and with GLP and cGMP regulations.
International companies frequently face this
kind of problem. Their laboratories have to comply with regulations
from different countries simultaneously with quality and
accreditation standards. The solution to this problem is to combine
all regulations and quality standards in a single quality manual and
a single set of operating procedures. The recommended documents and
how they relate to each other are shown in Figure 7.
The quality manual should place the company’s
own quality system first and foremost. This may be based on a
well-known laboratory quality standard, such as ISO/IEC 17025. The
quality manual and operating procedures should include aspects of
various regulations and quality standards applied within the
company. For specific requirements of single regulations, the
quality manual and procedures should include sections that only
apply to those particular regulations. For example, the
‘responsibility’ section would mention that for GLP studies, the
function of a study director is required. The tasks and
responsibilities should be described in an SOP.
Specific and/or Generic Procedures
Labcompliance procedures are designed to be
specific for ISO 17025 when it makes sense. Others are generic
enough such that they can be used for all quality system. To learn
more about procedures included the ISO 17025 accreditation package,
click here
- ISO/IEC 17025, General Requirements for
the Competence of Testing and Calibration Laboratories, 2005.
- Labcompliance, ISO/IEC17025 Accreditation
Package, 2009.
http://www.labcompliance.com/books/iso17025
- EURACHEM/CITAC Guide, Use of Uncertainty
Information in Compliance Assessment, 2007.
- EURACHEM/CITAC Guide CG4, Quantifying
Uncertainty in Analytical Measurement, ISBN 0-948926-15-5, 2000.
- EUROLAB, Measurement Uncertainty
Revisited: Alternative Approaches to Uncertainty Evaluation,
2007.
- ILAC, Introducing the Concept of
Uncertainty in Association with the Application of the Standard
ISO/IEC 17025, 2002.
- ILAC Policy on Traceability of
Measurement Results, 2002.
- EURACHEM/CITAC Guide, Traceability in
Chemical Measurement: A Guide to Achieving Comparable Results in
Chemical Measurement, 2003.
- EUROLAB, Guide to the Evaluation of
Measurement Uncertainty for Quantitative Test Results, 2006.
- ISO Guide 98-3, Uncertainty of
measurement -- Part 3: Guide to the Expression of Uncertainty in
Measurement (GUM:1995), Geneve, Switzerland, 2008.
- P. Bedson and M. Sargent, The Development
and Application of Guidance on Equipment Qualification of
Analytical Instruments, Accreditation and Quality Assurance, 1
(6), 265-274 (1996).
- EUROLAB Technical Report 2/2006,
Management of Computers and Software in Laboratories with
Reference to ISO/IEC 17025:2005.
- L. Huber, Validation and Qualification in
Analytical Laboratories, Interpharm, Informa Healthcare, New
York, USA, 1998. Second revision 2007.
- M. Thompson, S. Ellison and R. Wood,
International Harmonized Protocol for Proficiency Testing of
Chemical Analytical Laboratories, Pure Appl. Chem., Vol. 78, No.
1, pp. 145–196, 2006,
www.iupac.org/publications/pac/2006/pdf/7801x0145.pdf.
- American Association for Laboratory
Accreditation, A2LA Policy on Measurement Traceability, 2005.
- ISO/IEC Guide 2, Standardization and
Related Activities: General Vocabulary, 2005.
- X. R. Pan, Hierarchy of Reference
Materials Certified for Chemical Composition, Metrologia 34
(1997) 35-39.
To be updated on an ongoing basis
This chapter will be updated based on
questions and contributions from visitors
Q: Is ISO 17025 Accreditation required and
sufficient to comply with FDA Regulations, such as GLP and GMP?
Most requirements are the same or very
similar, but ISO 17025 is neither required nor sufficient. However,
many FDA laboratories have ISO 17025 implemented.
Rev 2: 2005
Q: When has ISO/IEC 17025 been introduced ?
Rev 1: 1999
Rev 2: 2005
Q: Is there any website where you can view or
download the ISO 17025 Standard?
ISO Standards are copyright protected. I am
not aware of a website with free download.
Q: Which organizations can award ISO 17025?
There are many in each country. To find them
in your area, use Google Search.
Q: What to do in case there is no proficiency
testing scheme for my application?
Do everything you can to proof accuracy of
your analytical data. Mechanisms include: extensive use of certified
reference material, verification of accuracy with an independent
analysis method and cross check of results within other departments
in you organization. Document the steps you took.
Q: Is an ISO/IEC 17025 accredited laboratory
also automatically ISO 9001 Certified?
ISO 17025 has a statement that testing and
calibration laboratories that comply with the International Standard
will also operate in accordance with ISO 9001.
On the other hand it states that
demonstrated conformity to this International Standard does not
imply conformity of the quality management system within which the
laboratory operates to all the requirements of ISO 9001. This means,
ISO 17025 conforms to related requirements of ISO 9001, but it does
not contain all the requirements as specified in ISO 90001. So the
answer to question above is 'no'.
This chapter will be updated on an on-going
basis